The Weighing Equipment (Automatic Catchweighing Instruments) Regulations (Northern Ireland) 2006

 Help about what version

What Version

 Help about advanced features

Advanced Features

Changes over time for: The Weighing Equipment (Automatic Catchweighing Instruments) Regulations (Northern Ireland) 2006 (Schedules only)

 Help about opening options

Alternative versions:

Changes to legislation:

There are outstanding changes not yet made by the legislation.gov.uk editorial team to The Weighing Equipment (Automatic Catchweighing Instruments) Regulations (Northern Ireland) 2006. Any changes that have already been made by the team appear in the content and are referenced with annotations. Help about Changes to Legislation

Close

Changes to Legislation

Revised legislation carried on this site may not be fully up to date. Changes and effects are recorded by our editorial team in lists which can be found in the ‘Changes to Legislation’ area. Where those effects have yet to be applied to the text of the legislation by the editorial team they are also listed alongside the legislation in the affected provisions. Use the ‘more’ link to open the changes and effects relevant to the provision you are viewing.

View outstanding changes

Changes and effects yet to be applied to :

Prospective

Regulations 2(1) and 6(c)

SCHEDULE 1N.I.(Accuracy Classes for Catchweighing Instruments)

Commencement Information

I1Sch. 1 in operation at 8.5.2006, see reg. 1

Instruments designated Class Y(a) and Y(b)N.I.

An instrument which is made in accordance with an approved pattern and which is marked Y(a) is a Class Y(a) instrument. The specifications for a Class Y(a) instrument are as follows—

Lower limit of the Minimum capacity “Min”:20 e
Number of scale intervals:≤10,000

An instrument which is made in accordance with an approved pattern and which is marked Y(b) is a Class Y(b) instrument. The specifications for a Class Y(b) instrument are as follows—

Lower limit of the Minimum capacity “Min”:10 e
Number of scale intervals:≤1,000

Regulation 5(a)

SCHEDULE 2N.I.(Manner of Erection and Installation)

Commencement Information

I2Sch. 2 in operation at 8.5.2006, see reg. 1

Installation (Principle from Clause 5.3.1 of OIML R51)N.I.

The installation of an automatic catchweighing instrument shall be so designed that an automatic weighing operation will be the same whether for the purposes of testing or for use for a transaction.

Cleaning and testingN.I.

Every automatic catchweighing instrument shall be positioned so as to facilitate cleaning and testing.

Dynamic setting (Extract from Clause 3.2.3 of OIML R51)N.I.

Instruments with dynamic setting shall have a facility for any access to dynamic setting to be automatically and non-erasably recorded.

Regulation 5(b)

SCHEDULE 3N.I.(Requirements in respect of Use and Manner of Use)

Commencement Information

I3Sch. 3 in operation at 8.5.2006, see reg. 1

Maximum and minimum capacities (Extract from Clause 2.4 of OIML R51)N.I.

Maximum capacity (Max) and minimum capacity (Min) shall be specified by the manufacturer.

The minimum capacity shall not be less than:

for class Y(a):20 e
for class Y(b):10 e
for postal scales:5 e

Temperature (Principle from Clause 2.9.1 of OIML R51)N.I.

Instruments shall comply with the appropriate metrological and technical requirements at temperatures from –10°C to +40°C. However, for special applications the limits of the temperature range may differ from those given above but such a range shall not be less than 30°C and shall be specified in the descriptive markings.

Where an automatic catchweighing instrument is marked with a temperature range it shall not be used for trade in temperatures outside that range.

Tilting (Principle from Clause 2.9.3 of OIML R51)N.I.

Instruments which are not intended for installation in a fixed position and which do not have a level indicator shall comply with the appropriate metrological and technical requirements when tilted by 5%.

Where an automatic catchweighing instrument is fitted with a level indicating device it shall enable the instrument to be set to a tilt of 1% or less.

Specified purpose or manner of useN.I.

Where an automatic catchweighing instrument is marked (in accordance with Schedule 4) with a mark which signifies the purpose or manner of use, it shall not be used for a purpose or in a manner which does not accord with that marking.

Instruments of Class Y(b) shall only be used for weighing ballast or waste or other goods in accordance with the particulars of the approved pattern.

Definition of wasteN.I.

For the purposes of the second paragraph of the preceding section of this Schedule (Specified purpose or manner of use), “waste” shall be construed in accordance with Article 2(2) of the Waste and Contaminated Land (Northern Ireland) Order 1997(1), provided that “waste” shall include any waste disposed of for reprocessing or recycling purposes but shall not include any radioactive waste as defined in section 2 of the Radioactive Substances Act 1993(2).

Zero-setting device (Extract from Clause 3.3.1 of OIML R51)N.I.

An automatic zero-setting device shall operate:

  • only when the stability criteria are fulfilled

  • sufficiently often to ensure that the zero is maintained within 0.5 e.

A non-automatic zero-setting device shall not be operable during automatic operation.

A semi-automatic zero-setting device shall function only when the stability criteria are fulfilled.

Zero-tracking device (Extract from Clause 3.3.2 of OIML R51)N.I.

A zero-tracking device shall operate only when:

  • the indication is at zero, or at a negative net value equivalent to gross zero, and

  • the stability criteria are fulfilled, and

  • the corrections are not more than 0.5 e/second.

Tare device (Extract from Clause 3.3.3 of OIML R51)N.I.

A semi-automatic or automatic tare device shall operate only when the stability criteria are fulfilled.

A non-automatic or semi-automatic tare device shall not be operable during automatic operation.

Suitability for use (Extract from Clause 3.1 of OIML R51)N.I.

An instrument shall be designed to suit the method of operation and the loads for which it is intended. It shall be of adequately robust construction in order that it maintains its metrological characteristics.

Minimum capacity (Extract from Clause 3.5 of OIML R51)N.I.

Printing a weighing result below minimum capacity shall not be possible.

Maximum capacity (Extract from Clause 3.4.3 of OIML R51)N.I.

There shall be no indication above Max + 9 e.

Rate of operationN.I.

An automatic catchweighing instrument shall not be used above the maximum rate of operation.

Regulation 6(b)

SCHEDULE 4N.I.(Descriptive Markings and Verification Marks: Principles from Part 1 of OIML R51 and Additional Descriptive Marking)

Descriptive markingsN.I.

3.8  Instruments shall bear the following markings.

Commencement Information

I4Sch. 4 para. 3.8 in operation at 8.5.2006, see reg. 1

Markings shown in fullN.I.

  • 3.8.1  name or identification mark of the manufacturer

  • name or identification mark of the importer (if applicable)

  • serial number and type designation of the instrument

  • maximum rate of operation (if applicable) in the form: … loads per minute

  • maximum speed of load transport system (if applicable) in the form: … m/s

  • electrical supply voltage in the form: … V

  • electrical supply frequency in the form: … Hz

  • working fluid pressure (if applicable) in the form: … kPa

  • adjustment range referred to set point (if applicable) in the form: ± … g or % (of set point value)

Commencement Information

I5Sch. 4 para. 3.8.1 in operation at 8.5.2006, see reg. 1

Markings shown in codeN.I.

3.8.2  

• pattern approval sign
• indication of the class of accuracy Y(y)
• verification scale interval in the form:e = …
• actual scale interval in the form:d = …
• maximum capacity in the form:Max = …
• minimum capacity in the form:Min = …
• maximum tare additive in the form:T = + …
• maximum tare subtractive in the form:T = − …

Commencement Information

I6Sch. 4 para. 3.8.2 in operation at 8.5.2006, see reg. 1

Supplementary markingsN.I.

3.8.3  Depending upon the particular use of the instrument, supplementary markings may be required on pattern approval by the metrological authority issuing the pattern approval certificate, (e.g. temperature range.)

Additional markings may be required on initial verification to specify types of packs and related weighing conditions.

Commencement Information

I7Sch. 4 para. 3.8.3 in operation at 8.5.2006, see reg. 1

Presentation of descriptive markingsN.I.

3.8.4  Descriptive markings shall be indelible and of a size, shape and clarity that permit legibility under normal conditions of use.

They shall be grouped together in a clearly visible place on the instrument, either on a descriptive plate fixed near the indicating device or on the indicating device itself. It shall be possible to seal the plate bearing the markings, unless it cannot be removed without being destroyed.

The descriptive markings may be shown on a programmable display which is controlled by software. In this case, means shall be provided for any access to reprogramming of the markings to be automatically and non-erasably recorded, e.g. by traceable access software. When a programmable display is used, the plate of the instrument shall bear at least the following markings:

  • type and designation of the instrument

  • name or identification mark of the manufacturer

  • pattern approval number

  • electrical supply voltage

  • electrical supply frequency

  • pneumatic pressure.

Commencement Information

I8Sch. 4 para. 3.8.4 in operation at 8.5.2006, see reg. 1

Verification marksN.I.

Commencement Information

I9Sch. 4 para. 3.9 in operation at 8.5.2006, see reg. 1

PositionN.I.

3.9.1  Instruments shall have a place for the application of verification marks. This place shall:

  • be such that the part on which it is located cannot be removed from the instrument without damaging the marks

  • allow easy application of the mark without changing the metrological qualities of the instrument

  • be visible without the instrument or its protective covers having to be moved when it is in service.

Commencement Information

I10Sch. 4 para. 3.9.1 in operation at 8.5.2006, see reg. 1

MountingN.I.

3.9.2  Instruments required to bear verification marks shall have a verification mark support, at the place provided for above, which shall ensure the conservation of the marks.

When the mark is made with a stamp the support may consist of a strip of lead or any other material with similar qualities, inserted into a plate fixed to the instrument, or in a cavity bored in the instrument.

Commencement Information

I11Sch. 4 para. 3.9.2 in operation at 8.5.2006, see reg. 1

Additional Descriptive MarkingN.I.

Automatic catchweighing instruments shall bear the additional descriptive marking “R51” which shall be presented in accordance with the provisions of clause 3.9.2 of Part 1 of OIML R51.

Regulation 10(3)

SCHEDULE 5N.I.(Prescribed Limits of Error)

Commencement Information

I12Sch. 5 in operation at 8.5.2006, see reg. 1

Maximum permissible errors for class Y(y) instrumentsN.I.

(Principle from Clause 2.3 of OIML R51)

The maximum permissible error for any load equal to or greater than the minimum capacity (Min) and equal to or less than the maximum capacity (Max) in automatic operation shall be as specified in Table 1. (Note that the mpe-value includes the digital rounding error of the indicating device.)

Table 1

Load (m) expressed in verification scale intervals (e)Maximum permissible error for class Y(y) instruments
Class Y(a)Class Y(b)Initial verificationIn-service
0 <m ≤ 5000 <m ≤50±1.5 e±2 e
500 <m ≤ 2,00050 <m ≤ 200±2 e±3 e
2,000 <m ≤ 10,000200 <m ≤ 1,000±2.5 e±4 e

(Principle from Clause 2.5.2 of OIML R51)

The maximum permissible error for any load equal to or greater than the minimum capacity (Min) and equal to or less than the maximum capacity (Max) for static weighing in non-automatic operation shall be as specified in Table 2.

Table 2

Load (m) expressed in verification scale intervals (e)Maximum permissible error for class Y(y) instruments
Class Y(a)Class Y(b)Initial verificationIn-service
0 <m ≤ 5000 <m ≤ 50±0.5 e±1 e
500 <m ≤ 2,00050 <m ≤ 200±1 e±2 e
2,000 <m ≤ 10,000200 <m ≤ 1,000±1.5 e±3 e

Back to top

Options/Help

Print Options

Close

Legislation is available in different versions:

Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.

Original (As Enacted or Made): The original version of the legislation as it stood when it was enacted or made. No changes have been applied to the text.

Close

See additional information alongside the content

Geographical Extent: Indicates the geographical area that this provision applies to. For further information see ‘Frequently Asked Questions’.

Show Timeline of Changes: See how this legislation has or could change over time. Turning this feature on will show extra navigation options to go to these specific points in time. Return to the latest available version by using the controls above in the What Version box.

Close

Opening Options

Different options to open legislation in order to view more content on screen at once

Close

More Resources

Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:

  • the original print PDF of the as enacted version that was used for the print copy
  • lists of changes made by and/or affecting this legislation item
  • confers power and blanket amendment details
  • all formats of all associated documents
  • correction slips
  • links to related legislation and further information resources
Close

Timeline of Changes

This timeline shows the different points in time where a change occurred. The dates will coincide with the earliest date on which the change (e.g an insertion, a repeal or a substitution) that was applied came into force. The first date in the timeline will usually be the earliest date when the provision came into force. In some cases the first date is 01/02/1991 (or for Northern Ireland legislation 01/01/2006). This date is our basedate. No versions before this date are available. For further information see the Editorial Practice Guide and Glossary under Help.

Close

More Resources

Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:

  • the original print PDF of the as made version that was used for the print copy
  • correction slips

Click 'View More' or select 'More Resources' tab for additional information including:

  • lists of changes made by and/or affecting this legislation item
  • confers power and blanket amendment details
  • all formats of all associated documents
  • links to related legislation and further information resources