Search Legislation

Directive 2014/32/EU of the European Parliament and of the CouncilShow full title

Directive 2014/32/EU of the European Parliament and of the Council of 26 February 2014 on the harmonisation of the laws of the Member States relating to the making available on the market of measuring instruments (recast) (Text with EEA relevance)

 Help about what version

What Version

 Help about UK-EU Regulation

Legislation originating from the EU

When the UK left the EU, legislation.gov.uk published EU legislation that had been published by the EU up to IP completion day (31 December 2020 11.00 p.m.). On legislation.gov.uk, these items of legislation are kept up-to-date with any amendments made by the UK since then.

Close

This item of legislation originated from the EU

Legislation.gov.uk publishes the UK version. EUR-Lex publishes the EU version. The EU Exit Web Archive holds a snapshot of EUR-Lex’s version from IP completion day (31 December 2020 11.00 p.m.).

Status:

This is the original version (as it was originally adopted).

SPECIFIC REQUIREMENTS

Instrument Classes

1.Two classes (0 and I) are being defined for exhaust gas analysers. The relevant minimum measuring ranges for these classes are shown in Table 1.
Table 1
Classes and measuring ranges
ParameterClasses 0 and I
CO fractionfrom 0 to 5 % vol
CO2 fractionfrom 0 to 16 % vol
HC fractionfrom 0 to 2 000 ppm vol
O2 fractionfrom 0 to 21 % vol
λfrom 0,8 to 1,2

Rated operating conditions

2.The values of the operating conditions shall be specified by the manufacturer as follows:
2.1.

For the climatic and mechanical influence quantities:

  • a minimum temperature range of 35 °C for the climatic environment;

  • the mechanical environment class that applies is M1.

2.2.

For the electrical power influence quantities:

  • the voltage and frequency range for the AC voltage supply;

  • the limits of the DC voltage supply.

2.3.

For the ambient pressure:

  • the minimum and the maximum values of the ambient pressure are for both classes: pmin ≤ 860 hPa, pmax ≥ 1 060 hPa.

Maximum permissible errors (MPEs)

3.The MPEs are defined as follows:
3.1.

For each of the fractions measured, the maximum error value permitted under rated operating conditions according to point 1.1 of Annex I is the greater of the two values shown in Table 2. Absolute values are expressed in % vol or ppm vol, percentage values are percent of the true value.

Table 2
MPEs
ParameterClass 0Class I
CO fraction

± 0,03 % vol

± 5 %

± 0,06 % vol

± 5 %

CO2 fraction

± 0,5 % vol

±5 %

± 0,5 % vol

± 5 %

HC fraction

± 10 ppm vol

± 5 %

± 12 ppm vol

± 5 %

O2 fraction

± 0,1 % vol

± 5 %

± 0,1 % vol

± 5 %

3.2.

The MPE on lambda calculation is 0,3 %. The conventional true value is calculated according to the formula set out in point 5.3.7.3 of Regulation No 83 of the Economic Commission for Europe of the United Nations (UN/ECE)(1).

For this purpose, the values displayed by the instrument are used for calculation.

Permissible effect of disturbances

4.For each of the volume fractions measured by the instrument, the critical change value is equal to the MPE for the parameter concerned.
5.The effect of an electromagnetic disturbance shall be such that:
  • either the change in the measurement result is not greater than the critical change value laid down in point 4;

  • or the presentation of the measurement result is such that it cannot be taken for a valid result.

Other requirements

6.The resolution shall be equal to or of one order of magnitude higher than the values shown in Table 3.
Table 3
a

0,01 % vol for measurand values below or equal to 4 % vol, otherwise 0,1 % vol.

Resolution
COCO2O2HC
Class 0 and class I0,01 % vol0,1 % vol a1 ppm vol

The lambda value shall be displayed with a resolution of 0,001.

7.The standard deviation of 20 measurements shall not be greater than one third of the modulus of the MPE for each applicable gas volume fraction.
8.For measuring CO, CO2 and HC, the instrument, including the specified gas handling system, must indicate 95 % of the final value as determined with calibration gases within 15 seconds after changing from a gas with zero content, e.g. fresh air. For measuring O2, the instrument under similar conditions must indicate a value differing less than 0,1 % vol from zero within 60 seconds after changing from fresh air to an oxygen-free gas.
9.The components in the exhaust gas, other than the components whose values are subject to the measurement, shall not affect the measurement results by more than the half of the modulus of the MPEs when those components are present in the following maximum volume fractions:
  • 6 % vol CO,

  • 16 % vol CO2,

  • 10 % vol O2,

  • 5 % vol H2,

  • 0,3 % vol NO,

  • 2 000 ppm vol HC (as n-hexane),

water vapor up to saturation.

10.An exhaust gas analyser shall have an adjustment facility that provides operations for zero-setting, gas calibration and internal adjustment. The adjustment facility for zero-setting and internal adjustment shall be automatic.
11.For automatic or semi-automatic adjustment facilities, the instrument shall be unable to make a measurement as long as the adjustments have not been made.
12.An exhaust gas analyser shall detect hydrocarbon residues in the gas handling system. It shall not be possible to carry out a measurement if the hydrocarbon residues, present before any measurement, exceed 20 ppm vol.
13.An exhaust gas analyser shall have a device for automatically recognising any malfunctioning of the sensor of the oxygen channel due to wear or a break in the connecting line.
14.If the exhaust gas analyser is capable to operate with different fuels (e.g. petrol or liquefied gas), there shall be the possibility to select the suitable coefficients for the Lambda calculation without ambiguity concerning the appropriate formula.

Back to top

Options/Help

Print Options

You have chosen to open the Whole Directive

The Whole Directive you have selected contains over 200 provisions and might take some time to download. You may also experience some issues with your browser, such as an alert box that a script is taking a long time to run.

Would you like to continue?

You have chosen to open Schedules only

The Schedules you have selected contains over 200 provisions and might take some time to download. You may also experience some issues with your browser, such as an alert box that a script is taking a long time to run.

Would you like to continue?

Close

Legislation is available in different versions:

Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.

Original (As adopted by EU): The original version of the legislation as it stood when it was first adopted in the EU. No changes have been applied to the text.

Close

Opening Options

Different options to open legislation in order to view more content on screen at once

Close

More Resources

Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:

  • the original print PDF of the as adopted version that was used for the EU Official Journal
  • lists of changes made by and/or affecting this legislation item
  • all formats of all associated documents
  • correction slips
  • links to related legislation and further information resources
Close

More Resources

Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:

  • the original print PDF of the as adopted version that was used for the print copy
  • correction slips

Click 'View More' or select 'More Resources' tab for additional information including:

  • lists of changes made by and/or affecting this legislation item
  • confers power and blanket amendment details
  • all formats of all associated documents
  • links to related legislation and further information resources