The manufacturer shall specify the rated operating conditions of the gas meter, taking into account:
Class | Qmax/Qmin | Qmax/Qt | Qr/Qmax |
---|---|---|---|
1,5 | ≥ 150 | ≥ 10 | 1,2 |
1,0 | ≥ 20 | ≥ 5 | 1,2 |
The gas meter shall be designed for the range of gases and supply pressures of the country of destination. In particular the manufacturer shall indicate:
the gas family or group;
the maximum operating pressure.
Class | 1,5 | 1,0 |
Qmin ≤ Q < Qt | 3 % | 2 % |
Qt ≤ Q ≤ Qmax | 1,5 % | 1 % |
The gas meter shall not exploit the MPEs or systematically favour any party.
the change in the measurement result is no greater than the critical change value as defined in point 3.1.3, or
the indication of the measurement result is such that it cannot be interpreted as a valid result, such as a momentary variation that cannot be interpreted, memorised or transmitted as a measuring result.
recover to operate within MPE, and
have all measurement functions safeguarded, and
allow recovery of all measurement data present just before the disturbance.
the quantity corresponding to half of the magnitude of the MPE in the upper zone on the measured volume;
the quantity corresponding to the MPE on the quantity corresponding to one minute at maximum flowrate.
Under installation conditions specified by the manufacturer, the effect of the flow disturbances shall not exceed one third of the MPE.
After an appropriate test, taking into account the period of time estimated by the manufacturer, has been performed, the following criteria shall be satisfied:
Metered quantity shall be displayed in cubic metre, or in kilogram.
The Whole Directive you have selected contains over 200 provisions and might take some time to download. You may also experience some issues with your browser, such as an alert box that a script is taking a long time to run.
Would you like to continue?
The Schedules you have selected contains over 200 provisions and might take some time to download. You may also experience some issues with your browser, such as an alert box that a script is taking a long time to run.
Would you like to continue?
Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.
Original (As adopted by EU): The original version of the legislation as it stood when it was first adopted in the EU. No changes have been applied to the text.
Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:
Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:
Click 'View More' or select 'More Resources' tab for additional information including: