Directive 2009/23/EC of the European Parliament and of the Council of 23 April 2009 on non-automatic weighing instruments (Codified version) (Text with EEA relevance) (repealed)
When the UK left the EU, legislation.gov.uk published EU legislation that had been published by the EU up to IP completion day (31 December 2020 11.00 p.m.). On legislation.gov.uk, these items of legislation are kept up-to-date with any amendments made by the UK since then.
Legislation.gov.uk publishes the UK version. EUR-Lex publishes the EU version. The EU Exit Web Archive holds a snapshot of EUR-Lex’s version from IP completion day (31 December 2020 11.00 p.m.).
This is the original version (as it was originally adopted).
The units of mass used shall be the legal units within the meaning of Council Directive 80/181/EEC of 20 December 1979 on the approximation of the laws of the Member States relating to units of measurement(1).
Subject to compliance with this condition, the following units are permitted:
SI units: kilogram, microgram, milligram, gram, tonne,
imperial unit: troy ounce, if weighing precious metals,
other non-SI unit: metric carat, if weighing precious stones.
For instruments that make use of the imperial unit of mass referred to above, the relevant essential requirements specified below shall be converted to that unit, using simple interpolation.
special
high
medium
ordinary
The specifications of these classes are given in Table 1.
Accuracy classes
The minimum capacity is reduced to 5 e for instruments in classes II and III for determining a conveying tariff.
1 × 10k, 2 × 10k, or 5 × 10k mass units,
k being any integer or zero.
d = e.
e = 1 × 10k g,
d < e ≤ 10 d,
except for instruments of class I with d < 10–4 g, for which e = 10–3 g.
Instruments equipped with an auxiliary indicating device shall belong to class I or class II. For these instruments the minimum capacity lower limits for these two classes are obtained from Table 1 by replacement in column 3 of the verification scale interval (e) by the actual scale interval (d).
If d < 10–4 g, the maximum capacity of class I may be less than 50 000 e.
Multiple weighing ranges are permitted, provided they are clearly indicated on the instrument. Each individual weighing range is classified according to point 3.1. If the weighing ranges fall into different accuracy classes the instrument shall comply with the severest of the requirements that apply for the accuracy classes in which the weighing ranges fall.
Multi-interval instruments shall not be equipped with an auxiliary indicating device.
where:
=
1, 2, … r,
=
partial weighing range number,
=
the total number of partial weighing ranges.
All capacities are capacities of net load, irrespective of the value of any tare used.
Multi-interval instruments
i = 1, 2, … r i = partial weighing range number r = total number of partial weighing ranges | |||||
a For i = r, the corresponding column of Table 1 applies, with e replaced by er. | |||||
Class | Verification scale interval (e) | Minimum capacity (Min) | Number of verification scale intervals | ||
---|---|---|---|---|---|
Minimum value | Minimum valuea | Maximum value | |||
I | 0,001 | g ≤ ei | 100 e1 | 50 000 | — |
II | 0,001 | g ≤ ei ≤ 0,05 g | 20 e1 | 5 000 | 100 000 |
0,1 | g ≤ ei | 50 e1 | 5 000 | 100 000 | |
III | 0,1 | g ≤ ei | 20 e1 | 500 | 10 000 |
IIII | 5 | g ≤ ei | 10 e1 | 50 | 1 000 |
The maximum permissible errors apply to the net value and tare value for all possible loads, excluding preset tare values.
Maximum permissible errors
Load | Maximum permissible error | |||
---|---|---|---|---|
Class I | Class II | Class III | Class IIII | |
0 ≤ m ≤ 50 000 e | 0 ≤ m ≤ 5 000 e | 0 ≤ m ≤ 500 e | 0 ≤ m ≤ 50 e | ±0,5 e |
50 000 e < m ≤ 200 000 e | 5 000 e < m ≤ 20 000 e | 500 e < m ≤ 2 000 e | 50 e < m ≤ 200 e | ±1,0 e |
200 000 e < m | 20 000 e < m ≤ 100 000 e | 2 000 e < m ≤ 10 000 e | 200 e < m ≤ 1 000 e | ±1,5 e |
The weighing results shall be sufficiently insensitive to changes in the position of the load on the load receptor.
5 °C for an instrument in class I,
15 °C for an instrument in class II,
30 °C for an instrument in class III or IIII.
In the absence of a manufacturer’s specification, the temperature range of – 10 °C to + 40 °C applies.
Instruments operated from battery power shall indicate whenever the voltage drops below the minimum required value and shall under those circumstances either continue to function correctly or be automatically put out of service.
Upon automatic detection of a significant fault, electronic instruments shall provide a visual or audible alarm that shall continue until the user takes corrective action or the fault disappears.
Digital electronic devices shall always exercise adequate control of the correct operation of the measuring process, of the indicating device, and of all data storage and data transfer.
Upon automatic detection of a significant durability error, electronic instruments shall provide a visual or audible alarm that shall continue until the user takes corrective action or the error disappears.
The indication of the weighing results and other weight values shall be accurate, unambiguous and non-misleading and the indicating device shall permit easy reading of the indication under normal conditions of use.
The names and symbols of the units referred to in point 1 of this Annex shall comply with the provisions of Directive 80/181/EEC with the addition of the symbol for the metric carat which shall be the symbol ‘ct’.
Indication shall be impossible above the maximum capacity (Max), increased by 9 e.
An auxiliary indicating device is permitted only to the right of the decimal mark. An extended indicating device may be used only temporarily, and printing shall be inhibited during its functioning.
Secondary indications may be shown, provided that they cannot be mistaken for primary indications.
Printed results shall be correct, suitably identified and unambiguous. The printing shall be clear, legible, non-erasable and durable.
When appropriate, instruments shall be fitted with a levelling device and a level indicator, sufficiently sensitive to allow proper installation.
Instruments may be equipped with zeroing devices. The operation of these devices shall result in accurate zeroing and shall not cause incorrect measuring results.
The instruments may have one or more tare devices and a preset tare device. The operation of the tare devices shall result in accurate zeroing and shall ensure correct net weighing. The operation of the preset tare device shall ensure correct determination of the calculated net value.
Instruments for direct sale to the public shall show all essential information about the weighing operation and, in the case of price-indicating instruments, shall clearly show the customer the price calculation of the product to be purchased.
The price to pay, if indicated, shall be accurate.
Price-computing instruments shall display the essential indications long enough for the customer to read them properly.
Price-computing instruments may perform functions other than per-article weighing and price computation only if all indications related to all transactions are printed clearly and unambiguously and are conveniently arranged on a ticket or label for the customer.
Instruments shall bear no characteristics that can cause, directly or indirectly, indications the interpretation of which is not easy or straightforward.
Instruments shall safeguard customers against incorrect sales transactions due to their malfunctioning.
Auxiliary indicating devices and extended indicating devices are not permitted.
Supplementary devices are permitted only if they cannot lead to fraudulent use.
Instruments similar to those normally used for direct sales to the public which do not satisfy the requirements of this Section must carry near to the display the indelible marking ‘Not to be used for direct sale to the public’.
Price labelling instruments shall meet the requirements of price indicating instruments for direct sale to the public, as far as applicable to the instrument in question. The printing of a price label shall be impossible below a minimum capacity.
Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.
Original (As adopted by EU): The original version of the legislation as it stood when it was first adopted in the EU. No changes have been applied to the text.
Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:
Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:
Click 'View More' or select 'More Resources' tab for additional information including: