Commission Directive 2005/10/EC of 4 February 2005 laying down the sampling methods and the methods of analysis for the official control of the levels of benzo(a)pyrene in foodstuffs (Text with EEA relevance) (repealed)
After exit day there will be three versions of this legislation to consult for different purposes. The legislation.gov.uk version is the version that applies in the UK. The EU Version currently on EUR-lex is the version that currently applies in the EU i.e you may need this if you operate a business in the EU.
The web archive version is the official version of this legislation item as it stood on exit day before being published to legislation.gov.uk and any subsequent UK changes and effects applied. The web archive also captured associated case law and other language formats from EUR-Lex.
This is the original version (as it was originally adopted).
Where no specific methods for the determination of benzo(a)pyrene in food are prescribed at Community level, laboratories may select any validated method provided the selected method meets the performance criteria indicated in the Table. The validation should ideally include a certified reference material.
Performance criteria for methods of analysis for benzo(a)pyrene
Parameter | Value/comment |
---|---|
Applicability | Food specified in Regulation (EC) No …/2005 |
Detection limit | No more than 0,3 μg/kg |
Limit of quantification | No more than 0,9 μg/kg |
Precision | HORRATr or HORRATR values of less than 1.5 in the validation collaborative trial |
Recovery | 50 %-120 % |
Specificity | Free from matrix or spectral interferences, verification of positive detection |
However, an uncertainty approach may also be used to assess the suitability of the method of analysis to be used by the laboratory. The laboratory may use a method which will produce results within a maximum standard uncertainty. The maximum standard uncertainty can be calculated using the following formula:
where:
is the maximum standard uncertainty
is the limit of detection of the method
is the concentration of interest
If an analytical method provides results with uncertainty measurements less than the maximum standard uncertainty the method will be equally suitable to one which meets the performance characteristics given in the Table.
Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.
Original (As adopted by EU): The original version of the legislation as it stood when it was first adopted in the EU. No changes have been applied to the text.
Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:
Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:
Click 'View More' or select 'More Resources' tab for additional information including: