Search Legislation

Commission Decision of 27 November 2009 amending Decision 2002/364/EC on common technical specifications for in vitro diagnostic medical devices (notified under document C(2009) 9464) (Text with EEA relevance) (2009/886/EC)

 Help about what version

What Version

 Help about advanced features

Advanced Features

Close

This is a legislation item that originated from the EU

After exit day there will be three versions of this legislation to consult for different purposes. The legislation.gov.uk version is the version that applies in the UK. The EU Version currently on EUR-lex is the version that currently applies in the EU i.e you may need this if you operate a business in the EU.

The web archive version is the official version of this legislation item as it stood on exit day before being published to legislation.gov.uk and any subsequent UK changes and effects applied. The web archive also captured associated case law and other language formats from EUR-Lex.

Changes to legislation:

There are currently no known outstanding effects for the Commission Decision of 27 November 2009 amending Decision 2002/364/EC on common technical specifications for in vitro diagnostic medical devices (notified under document C(2009) 9464) (Text with EEA relevance) (2009/886/EC), ANNEX. Help about Changes to Legislation

Close

Changes to Legislation

Revised legislation carried on this site may not be fully up to date. At the current time any known changes or effects made by subsequent legislation have been applied to the text of the legislation you are viewing by the editorial team. Please see ‘Frequently Asked Questions’ for details regarding the timescales for which new effects are identified and recorded on this site.

ANNEXU.K.

ANNEXU.K. COMMON TECHNICAL SPECIFICATIONS (CTS) FOR IN VITRO DIAGNOSTIC MEDICAL DEVICES

1.SCOPE

The common technical specifications set out in this Annex shall apply for the purposes of Annex II List A to Directive 98/79/EC.

2.DEFINITIONS AND TERMS

(Diagnostic) sensitivity

The probability that the device gives a positive result in the presence of the target marker.

True positive

A specimen known to be positive for the target marker and correctly classified by the device.

False negative

A specimen known to be positive for the target marker and misclassified by the device.

(Diagnostic) specificity

The probability that the device gives a negative result in the absence of the target marker.

False positive

A specimen known to be negative for the target marker and misclassified by the device.

True negative

A specimen known to be negative for the target marker and correctly classified by the device.

Analytical sensitivity

Analytical sensitivity may be expressed as the limit of detection, i.e. the smallest amount of the target marker that can be precisely detected.

Analytical specificity

Analytical specificity means the ability of the method to determine solely the target marker.

Nucleic acid amplification techniques (NAT)

The term “NAT” is used for tests for the detection and/or quantification of nucleic acids by either amplification of a target sequence, by amplification of a signal or by hybridisation.

Rapid test

“Rapid test” means qualitative or semi-quantitative in vitro diagnostic medical devices, used singly or in a small series, which involve non-automated procedures and have been designed to give a fast result.

Robustness

The robustness of an analytical procedure means the capacity of an analytical procedure to remain unaffected by small but deliberate variations in method parameters and provides an indication of its reliability during normal usage.

Whole system failure rate

The whole system failure rate means the frequency of failures when the entire process is performed as prescribed by the manufacturer.

Confirmation assay

Confirmation assay means an assay used for the confirmation of a reactive result from a screening assay.

Virus typing assay

Virus typing assay means an assay used for typing with already known positive samples, not used for primary diagnosis of infection or for screening.

Sero-conversion HIV samples

Sero-conversion HIV samples mean:

  • p24 antigen and/or HIV RNA positive, and

  • recognised by all of the antibody screening tests, and

  • positive or indeterminate confirmatory assays.

Early sero-conversion HIV samples

Early seroconversion HIV samples mean:

  • p24 antigen and/or HIV RNA positive, and

  • not recognised by all of the antibody screening tests, and

  • indeterminate or negative confirmatory assays.

3.COMMON TECHNICAL SPECIFICATIONS (CTS) FOR PRODUCTS REFERRED TO IN ANNEX II, LIST A OF DIRECTIVE 98/79/EC

3.1. CTS for performance evaluation of reagents and reagent products for the detection, confirmation and quantification in human specimens of markers of HIV infection (HIV 1 and 2), HTLV I and II, and hepatitis B, C, D
General principles
3.1.1.Devices which detect virus infections placed on the market for use as either screening or diagnostic tests, shall meet the requirements for sensitivity and specificity set out in Table 1. See also principle 3.1.11 for screening assays.
3.1.2.Devices intended by the manufacturer for testing body fluids other than serum or plasma, e.g. urine, saliva, etc., shall meet the same CTS requirements for sensitivity and specificity as serum or plasma tests. The performance evaluation shall test samples from the same individuals in both the tests to be approved and in a respective serum or plasma assay.
3.1.3.Devices intended by the manufacturer for self-test, i.e. home use, shall meet the same CTS requirements for sensitivity and specificity as respective devices for professional use. Relevant parts of the performance evaluation shall be carried out (or repeated) by appropriate lay users to validate the operation of the device and the instructions for use.
3.1.4.All performance evaluations shall be carried out in direct comparison with an established state-of-the-art device. The device used for comparison shall be one bearing CE marking, if on the market at the time of the performance evaluation.
3.1.5.If discrepant test results are identified as part of an evaluation, these results shall be resolved as far as possible, for example:
  • by evaluation of the discrepant sample in further test systems,

  • by use of an alternative method or marker,

  • by a review of the clinical status and diagnosis of the patient, and

  • by the testing of follow-up-samples.

3.1.6.Performance evaluations shall be performed on a population equivalent to the European population.
3.1.7.Positive specimens used in the performance evaluation shall be selected to reflect different stages of the respective disease(s), different antibody patterns, different genotypes, different subtypes, mutants, etc.
3.1.8.Sensitivity with true positives and sero-conversion samples shall be evaluated as follows:
3.1.8.1.

Diagnostic test sensitivity during sero-conversion has to represent the state of the art. Whether further testing of the same or additional sero-conversion panels is conducted by the notified body or by the manufacturer the results shall confirm the initial performance evaluation data (see Table 1). Sero-conversion panels should start with a negative bleed(s) and should have narrow bleeding intervals.

3.1.8.2.

For blood screening devices (with the exception of HBsAg and anti-HBc tests), all true positive samples shall be identified as positive by the device to be CE marked (Table 1). For HBsAg and anti-HBc tests the new device shall have an overall performance at least equivalent to that of the established device (see 3.1.4).

3.1.8.3.

Regarding HIV tests:

  • all sero-conversion HIV samples shall be identified as positive, and

  • at least 40 early sero-conversion HIV samples shall be tested. Results should conform to the state of the art.

3.1.9.Performance evaluation of screening assays shall include 25 positive (if available in the case of rare infections) “same day” fresh serum and/or plasma samples (≤ 1 day after sampling).
3.1.10.Negative specimens used in a performance evaluation shall be defined so as to reflect the target population for which the test is intended, for example blood donors, hospitalised patients, pregnant women, etc.
3.1.11.For performance evaluations for screening assays (Table 1) blood donor populations shall be investigated from at least two blood donation centres and consist of consecutive blood donations, which have not been selected to exclude first time donors.
3.1.12.Devices shall have a specificity of at least 99,5 % on blood donations, unless otherwise indicated in the accompanying tables. Specificity shall be calculated using the frequency of repeatedly reactive (i.e. false positive) results in blood donors negative for the target marker.
3.1.13.Devices shall be evaluated to establish the effect of potential interfering substances, as part of the performance evaluation. The potential interfering substances to be evaluated will depend to some extent on the composition of the reagent and configuration of the assay. Potential interfering substances shall be identified as part of the risk analysis required by the essential requirements for each new device but may include, for example:
  • specimens representing “related” infections,

  • specimens from multipara, i.e. women who have had more than one pregnancy, or rheumatoid factor positive patients,

  • for recombinant antigens, human antibodies to components of the expression system, for example anti-E. coli, or anti-yeast.

3.1.14.For devices intended by the manufacturer to be used with serum and plasma the performance evaluation must demonstrate serum to plasma equivalency. This shall be demonstrated for at least 50 donations (25 positive and 25 negative).
3.1.15.For devices intended for use with plasma the performance evaluation shall verify the performance of the device using all anticoagulants which the manufacturer indicates for use with the device. This shall be demonstrated for at least 50 donations (25 positive and 25 negative).
3.1.16.As part of the required risk analysis the whole system failure rate leading to false-negative results shall be determined in repeat assays on low-positive specimens.
3.1.17.If a new in vitro diagnostic medical device belonging to Annex II List A is not specifically covered by the common technical specification, the common technical specification for a related device should be taken into account. Related devices may be identified on different grounds, e.g. by the same or similar intended use or by similar risks.
3.2. Additional requirements for HIV antibody/antigen combined tests
3.2.1.HIV antibody/antigen combined tests intended for anti-HIV and p24 antigen detection which include claims for single p24 antigen detection shall follow Table 1 and Table 5, including criteria for analytical sensitivity for p24 antigen.
3.2.2.HIV antibody/antigen combined tests intended for anti-HIV and p24 detection which do not include claims for single p24 detection shall follow Table 1 and Table 5, excluding criteria for analytical sensitivity for p24.
3.3. Additional requirements for nucleic acid amplification techniques (NAT)

The performance evaluation criteria for NAT assays can be found in Table 2.

3.3.1.For target sequence amplification assays, a functionality control for each test sample (internal control) shall reflect the state of the art. This control shall as far as possible be used throughout the whole process, i.e. extraction, amplification/hybridisation, detection.
3.3.2.The analytical sensitivity or detection limit for NAT assays shall be expressed by the 95 % positive cut-off value. This is the analyte concentration where 95 % of test runs give positive results following serial dilutions of an international reference material for example a WHO standard or calibrated reference material.
3.3.3.Genotype detection shall be demonstrated by appropriate primer or probe design validation and shall also be validated by testing characterised genotyped samples.
3.3.4.Results of quantitative NAT assays shall be traceable to international standards or calibrated reference materials, if available, and be expressed in international units utilised in the specific field of application.
3.3.5.NAT assays may be used to detect virus in antibody negative samples, i.e. pre-sero-conversion samples. Viruses within immune-complexes may behave differently in comparison to free viruses, for example during a centrifugation step. It is therefore important that during robustness studies, antibody-negative (pre-sero-conversion) samples are included.
3.3.6.For investigation of potential carry-over, at least five runs with alternating high-positive and negative specimens shall be performed during robustness studies. The high positive samples shall comprise samples with naturally occurring high virus titres.
3.3.7.The whole system failure rate leading to false-negative results shall be determined by testing low-positive specimens. Low-positive specimens shall contain a virus concentration equivalent to three times the 95 % positive cut-off virus concentration.
3.4. CTS for the manufacturer’s release testing of reagents and reagent products for the detection, confirmation and quantification in human specimens of markers of HIV infection (HIV 1 and 2), HTLV I and II, and hepatitis B, C, D (immunological assays only)
3.4.1.The manufacturer’s release testing criteria shall ensure that every batch consistently identifies the relevant antigens, epitopes, and antibodies.
3.4.2.The manufacturer’s batch release testing for screening assays shall include at least 100 specimens negative for the relevant analyte.
3.5. CTS for performance evaluation of reagents and reagent products for determining the following blood group antigens: ABO blood group system ABO1 (A), ABO2 (B), ABO3 (A,B); Rh blood group system RH1 (D), RH2 (C), RH3 (E), RH4 (c), RH5 (e); Kell blood group system KEL1 (K)

Criteria for performance evaluation of reagents and reagent products for determining the blood groups antigens: ABO blood group system ABO1 (A), ABO2 (B), ABO3 (A,B); Rh blood group system RH1 (D), RH2 (C), RH3 (E), RH4 (c), RH5 (e); Kell blood group system KEL1 (K) can be found in Table 9.

3.5.1.All performance evaluations shall be carried out in direct comparison with an established state-of-the-art device. The device used for comparison shall be one bearing CE marking, if on the market at the time of the performance evaluation.
3.5.2.If discrepant test results are identified as part of an evaluation, these results shall be resolved as far as possible, for example:
  • by evaluation of the discrepant sample in further test systems,

  • by use of an alternative method,

3.5.3.Performance evaluations shall be performed on a population equivalent to the European population.
3.5.4.Positive specimens used in the performance evaluation shall be selected to reflect variant and weak antigen expression.
3.5.5.Devices shall be evaluated to establish the effect of potential interfering substances, as part of the performance evaluation. The potential interfering substances to be evaluated will depend to some extent on the composition of the reagent and configuration of the assay. Potential interfering substances shall be identified as part of the risk analysis required by the essential requirements for each new device.
3.5.6.For devices intended for use with plasma the performance evaluation shall verify the performance of the device using all anticoagulants which the manufacturer indicates for use with the device. This shall be demonstrated for at least 50 donations.
3.6. CTS for the manufacturer’s release testing of reagents and reagent products for determining the blood group antigens: ABO blood group system ABO1 (A), ABO2 (B), ABO3 (A,B); Rh blood group system RH1 (D), RH2 (C), RH3 (E), RH4 (c), RH5 (e); Kell blood group system KEL1 (K)
3.6.1.The manufacturer’s release testing criteria shall ensure that every batch consistently identifies the relevant antigens, epitopes, and antibodies.
3.6.2.Requirements for manufacturers batch release testing are outlined in Table 10.
Table 1

“Screening” assays: anti-HIV 1 and 2, anti-HTLV I and II, anti-HCV, HBsAg, anti-HBc

Anti-HIV-1/2Anti-HTLV-I/IIAnti-HCVHBsAgAnti-HBc
Diagnostic sensitivityPositive specimens

400 HIV-1

100 HIV-2

including 40 non-B subtypes, all available HIV/1 subtypes should be represented by at least 3 samples per subtype

300 HTLV-I

100 HTLV-II

400 (positive samples)

Including samples from different stages of infection and reflecting different antibody patterns.

Genotype 1-4: > 20 samples per genotype (including non-a subtypes of genotype 4);

5: > 5 samples;

6: if available

400

Including subtypeconsideration

400

Including evaluation of other HBV-markers

Sero-conversion panels

20 panels

10 further panels (at Notified Body or manufacturer)

To be defined when available

20 panels

10 further panels (at Notified Body or manufacturer)

20 panels

10 further panels (at Notified Body or manufacturer)

To be defined when available
Analytical sensitivityStandards0,130 IU/ml (Second International Standard for HBsAg, subtype adw2, genotype A, NIBSC code: 00/588)
SpecificityUnselected donors (including first-time donors)5 0005 0005 0005 0005 000
Hospitalised patients200200200200200
Potentially cross-reacting blood-specimens (RF+, related viruses, pregnant women, etc.)100100100100100
Table 2

NAT assays for HIV1, HCV, HBV, HTLV I/II (qualitative and quantitative; not molecular typing)

a

European Pharmacopoeia guideline.

Notes: Acceptance criteria for “whole system failure rate leading to false-neg results” is 99/100 assays positive.

For quantitative NATs a study shall be performed on at least 100 positive specimens reflecting the routine conditions of users (e.g. no pre-selection of specimens). Comparative results with another NAT test system shall be generated in parallel.

For qualitative NATs a study on diagnostic sensitivity shall be performed using at least 10 sero-conversion panels. Comparative results with another NAT test system shall be generated in parallel.

HIV1HCVHBVHTLV I/IIAcceptance criteria
NATqualitativequantitativequalitativequantitativequalitativequantitativequalitativequantitative
As for HIV quantitativeAs for HIV quantitativeAs for HIV quantitative

Sensitivity

Detection limit

Detection of analytical sensitivity (IU/ml; defined on WHO standards or calibrated reference materials)

According to EP validation guidelinea: several dilution series into borderline concentration; statistical analysis (e.g. Probit analysis) on the basis of at least 24 replicates; calculation of 95 % cut-off valueDetection limit: as for qualitative tests; Quantification limit: dilutions (half-log10 or less) of calibrated reference preparations, definition of lower, upper quantification limit, precision, accuracy, “linear” measuring range, “dynamic range”. Reproducibility at different concentration levels to be shownAccording to EP validation guidelinea: several dilution series into borderline concentration; statistical analysis (e.g. Probit analysis) on the basis of at least 24 replicates; calculation of 95 % cut-off valueAccording to EP validation guidelinea: several dilution series into borderline concentration; statistical analysis (e.g. Probit analysis) on the basis of at least 24 replicates; calculation of 95 % cut-off valueAccording to EP validation guidelinea: several dilution series into borderline concentration; statistical analysis (e.g. Probit analysis) on the basis of at least 24 replicates; calculation of 95 % cut-off value
Genotype/subtype detection/quantification efficiencyAt least 10 samples per subtype (as far as available)Dilution series of all relevant genotypes/subtypes, preferably of reference materials, as far as availableAt least 10 samples per genotype (as far as available)As far as calibrated genotype reference materials are availableAs far as calibrated genotype reference materials are available
Cell culture supernatants (could substitute for rare HIV-1 subtypes)Transcripts or plasmids quantified by appropriate methods may be used.
According to EP validation guidelineaas far as calibrated subtype reference materials are available; in vitro transcripts could be an optionAccording to EP validation guidelineaas far as calibrated subtype reference materials are available; in vitro transcripts could be an optionAccording to EP validation guidelineaas far as calibrated subtype reference materials are available; in vitro transcripts could be an optionAccording to EP validation guidelineaas far as calibrated subtype reference materials are available; in vitro transcripts could be an option
Diagnostic specificity negative samples500 blood donors100 blood donors500 blood donors500 blood donors500 individual blood donations
Potential cross-reactive markersBy suitable assay design evidence (e.g. sequence comparison) and/or testing of at least 10 human retrovirus (e.g. HTLV)-positive samplesAs for qualitative testsBy assays design and/or testing of at least 10 human flavivirus (e.g. HGV, YFV) positive samplesBy assays design and/or testing of at least 10 other DNA-virus positive samplesBy assay design and/or testing of at least 10 human retrovirus (e.g. HIV-) positive samples
RobustnessAs for qualitative tests
Cross-contaminationAt least 5 runs using alternating high positive (known to occur naturally) and negative samplesAt least 5 runs using alternating high positive (known to occur naturally) and negative samplesAt least 5 runs using alternating high positive (known to occur naturally) and negative samplesAt least 5 runs using alternating high positive (known to occur naturally) and negative samples
InhibitionInternal control preferably to go through the whole NAT procedureInternal control preferably to go through the whole NAT procedureInternal control preferably to go through the whole NAT procedureInternal control preferably to go through the whole NAT procedure
Whole system failure rate leading to false-neg resultsAt least 100 samples virus-spiked with 3 × the 95 % pos cut-off concentrationAt least 100 samples virus-spiked with 3 × the 95 % pos cut-off concentrationAt least 100 samples virus-spiked with 3 × the 95 % pos cut-off concentrationAt least 100 samples virus-spiked with 3 × the 95 % pos cut-off concentration99/100 assays positive
Table 3

Rapid tests: anti-HIV 1 and 2, anti-HCV, HBsAg, anti-HBc, anti-HTLV I and II

Anti-HIV 1/2Anti-HCVHBsAgAnti-HBcAnti-HTLV I/IIAcceptance criteria
Diagnostic sensitivityPositive specimensSame criteria as for screening assaysSame criteria as for screening assaysSame criteria as for screening assaysSame criteria as for screening assaysSame criteria as for screening assaysSame criteria as for screening assays
Sero-conversion panelsSame criteria as for screening assaysSame criteria as for screening assaysSame criteria as for screening assaysSame criteria as for screening assaysSame criteria as for screening assaysSame criteria as for screening assays
Diagnostic specificityNegative specimens1 000 blood donations1 000 blood donations1 000 blood donations1 000 blood donations1 000 blood donations?>≥ 99 % (anti-HBc: ≥ 96 %)
200 clinical specimens200 clinical specimens200 clinical specimens200 clinical specimens200 clinical specimens
200 samples from pregnant women200 samples from pregnant women200 samples from pregnant women200 samples from pregnant women
100 potentially interfering samples100 potentially interfering samples100 potentially interfering samples100 potentially interfering samples100 potentially interfering samples
Table 4

Confirmatory/supplementary assays for anti-HIV 1 and 2, anti-HTLV I and II, anti-HCV, HBsAg

a

Acceptance criteria no neutralisation for HBsAg confirmatory assay.

Anti-HIV confirmatory assayAnti-HTLV confirmatory assayHCV supplementary assayHBsAg confirmatory assayAcceptance criteria
Diagnostic sensitivityPositive specimens200 HIV-1 and 100 HIV-2200 HTLV-I and 100 HTLV-II300 HCV (positive samples)300 HBsAgCorrect identification as positive (or indeterminate), not negative
Including samples from different stages of infection and reflecting different antibody patterns

Including samples from different stages of infection and reflecting different antibody patterns.

Genotypes 1 – 4: > 20 samples (including non-a subtypes of genotype 4);

5: > 5 samples;

6: if available

Including samples from different stages of infection

20 “high pos” samples (> 26 IU/ml); 20 samples in the cut-off range

Sero-conversion panels15 sero-conversion panels/low titre panels15 sero-conversion panels/low titre panels15 sero-conversion panels/low titre panels
Analytical sensitivityStandardsSecond International Standard for HBsAg, subtype adw2, genotype A, NIBSC code: 00/588
Diagnostic specificityNegative specimens200 blood donations200 blood donation200 blood donations10 false positives as available from the performance evaluation of the screening assaya.No false-positive results/a no neutralisation
200 clinical samples including pregnant women200 clinical samples including pregnant women200 clinical samples including pregnant women
50 potentially interfering samples, including samples with indeterminate results in other confirmatory assays50 potentially interfering samples including samples with indeterminate results in other confirmatory assays50 potentially interfering samples including samples with indeterminate results in other supplementary assays50 potentially interfering samples
Table 5

HIV 1 antigen

HIV-1 antigen assayAcceptance criteria
Diagnostic sensitivityPositive specimens

50 HIV-1 Ag-positive

50 cell culture supernatants including different HIV-1 subtypes and HIV-2

Correct identification (after neutralisation)
Sero-conversion panels20 sero-conversion panels/low titre panels
Analytical sensitivityStandardsHIV-1 p24 Antigen, First International Reference Reagent, NIBSC code: 90/636≤ 2 IU/ml
Diagnostic specificity

200 blood donations

200 clinical samples

50 potentially interfering samples

≥ 99,5 % after neutralisation
Table 6

Serotyping and genotyping assay: HCV

HCV serotyping and genotyping assayAcceptance criteria
Diagnostic sensitivityPositive specimens

200 (positive samples)

Including samples from different stages of infection and reflecting different antibody patterns.

Genotypes 1 – 4: > 20 samples (including non-a subtypes of genotype 4);

5: > 5 samples;

6: if available

≥ 95 % agreement between serotyping and genotyping

[X1> 95 % agreement between genotyping and sequencing]

Diagnostic specificityNegative specimens100
Table 7

HBV markers: anti-HBs, anti HBc IgM, anti-HBe, HBeAg

Anti-HBsAnti-HBc IgMAnti-HBeHBeAgAcceptance criteria
Diagnostic sensitivityPositive specimens100 vaccinees200200200≥ 98 %
100 naturally infected persons

Including samples from different stages of infection (acute/chronic, etc.)

The acceptance criteria should only be applied on samples from acute infection stage.

Including samples from different stages of infection (acute/chronic, etc.)Including samples from different stages of infection (acute/chronic, etc.)
Sero-conversion panels10 follow-ups or anti-HBs sero-conversionsWhen available
Analytical sensitivityStandardsWHO First International Reference Preparation 1977; NIBSC, United KingdomHBe — Referenzantigen 82; PEI GermanyAnti-HBs: < 10 mIU/ml
Diagnostic specificityNegative specimens500200 blood donations200 blood donation200 blood donations≥ 98 %
Including clinical samples200 clinical samples200 clinical samples200 clinical samples
50 potentially interfering samples50 potentially interfering samples50 potentially interfering samples50 potentially interfering samples
Table 8

HDV markers: anti-HDV, anti-HDV IgM, delta antigen

Anti-HDVAnti-HDV IgMDelta antigenAcceptance criteria
Diagnostic sensitivityPositive specimens1005010≥ 98 %
Specifying HBV markersSpecifying HBV markersSpecifying HBV markers
Diagnostic specificityNegative specimens200200200≥ 98 %
Including clinical samplesIncluding clinical samplesIncluding clinical samples
50 potentially interfering samples50 potentially interfering samples50 potentially interfering samples
Table 9

Blood group antigens in the ABO, Rh and Kell blood group systems

123
SpecificityNumber of tests per recommended methodTotal number of samples to be tested for a launch productTotal number of samples to be tested for a new formulation, or use of well-characterised reagents
Anti-ABO1 (anti-A), anti-ABO2 (anti-B), anti-ABO3 (anti-A,B)5003 0001 000
Anti-RH1 (anti-D)5003 0001 000
Anti-RH2 (anti-C), anti-RH4 (anti-c), anti-RH3 (anti-E)1001 000200
Anti-RH5 (anti-e)100500200
Anti-KEL1 (anti-K)100500200
Acceptance criteria:

All of the above reagents shall show comparable test results with established reagents with acceptable performance with regard to claimed reactivity of the device. For established reagents, where the application or use has been changed or extended, further testing should be carried out in accordance with the requirements outlined in column 1 (above).

Performance evaluation of anti-D reagents shall include tests against a range of weak RH1 (D) and partial RH1 (D) samples, depending on the intended use of the product.

Qualifications:
Clinical samples

:

10 % of the test population

Neonatal specimens

:

> 2 % of the test population

ABO samples

:

> 40 % A, B positives

“weak D”

:

> 2 % of RH1 (D) positives

Table 10 Batch release criteria for reagents and reagent products to determine blood group antigens in the ABO, Rh and Kell blood group systems Specificity testing requirements on each reagent
1. Test reagents
a

Only by recommended techniques where reactivity against these antigens is claimed.

Note: Polyclonal reagents must be tested against a wider panel of cells to confirm specificity and exclude presence of unwanted contaminating antibodies.

Blood group reagentsMinimum number of control cells to be tested
Positive reactionsNegative reactions
A1A2BAxB0
Anti-ABO1 (anti-A)222a22
BA1BA10
Anti-ABO2 (anti-B)2222
A1A2AxB0
Anti-ABO3 (anti-A,B)22224
R1rR2rWeakDr’rr’rrr
Anti-RH1 (anti-D)222a111
R1R2R1rr’rR2R2r’rrr
Anti-RH2 (anti-C)211111
R1R2R1rr’rR1R1
Anti-RH4 (anti-c)1213
R1R2R2rr’rR1R1r’rrr
Anti-RH 3 (anti-E)211111
R1R2R2rr’rR2R2
Anti-RH5 (anti-e)2113
Kkkk
Anti-KEL1 (anti-K)43
Acceptance criteria:

Each batch of reagent must exhibit unequivocal positive or negative results by all recommended techniques in accordance with the results obtained from the performance evaluation data.

2. Control materials (red cells)

The phenotype of red cells used in the control of blood typing reagents listed above should be confirmed using established device.

Back to top

Options/Help

Print Options

Close

Legislation is available in different versions:

Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.

Original (As adopted by EU): The original version of the legislation as it stood when it was first adopted in the EU. No changes have been applied to the text.

Close

See additional information alongside the content

Geographical Extent: Indicates the geographical area that this provision applies to. For further information see ‘Frequently Asked Questions’.

Show Timeline of Changes: See how this legislation has or could change over time. Turning this feature on will show extra navigation options to go to these specific points in time. Return to the latest available version by using the controls above in the What Version box.

Close

Opening Options

Different options to open legislation in order to view more content on screen at once

Close

More Resources

Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:

  • the original print PDF of the as adopted version that was used for the EU Official Journal
  • lists of changes made by and/or affecting this legislation item
  • all formats of all associated documents
  • correction slips
  • links to related legislation and further information resources
Close

Timeline of Changes

This timeline shows the different versions taken from EUR-Lex before exit day and during the implementation period as well as any subsequent versions created after the implementation period as a result of changes made by UK legislation.

The dates for the EU versions are taken from the document dates on EUR-Lex and may not always coincide with when the changes came into force for the document.

For any versions created after the implementation period as a result of changes made by UK legislation the date will coincide with the earliest date on which the change (e.g an insertion, a repeal or a substitution) that was applied came into force. For further information see our guide to revised legislation on Understanding Legislation.

Close

More Resources

Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:

  • the original print PDF of the as adopted version that was used for the print copy
  • correction slips

Click 'View More' or select 'More Resources' tab for additional information including:

  • lists of changes made by and/or affecting this legislation item
  • confers power and blanket amendment details
  • all formats of all associated documents
  • links to related legislation and further information resources