Regulation 2(1)
Regulation 3(1)
1.—(1) The device shall be designed and manufactured in such a way that, when implanted under the conditions and for the purposes laid down by the manufacturer, its use does not compromise the clinical condition or the safety of the patient.
(2) The device shall not present any foreseeable risk to the person implanting it or to other persons.
2. The device must achieve the performances intended by its manufacturer and, in particular, must be designed and manufactured in such a way that it is suitable for one or more of the following functions as specified by him:—
(a)diagnosis, prevention, monitoring, treatment or alleviation of disease or injury;
(b)investigation, replacement or modification of the anatomy or of a physiological process;
(c)control of conception.
3. The characteristics and performances referred to in paragraphs 1 and 2 shall not be adversely affected to such a degree that the clinical condition or safety of the patient or of other persons is compromised during the lifetime of the device expected by the manufacturer, where the device is subjected to stresses which may occur during normal conditions of use.
4. The device shall be designed, manufactured and packed in such a way that its characteristics and performances are not adversely affected if the conditions as to storage and transport laid down by the manufacturer, such as conditions relating to temperature or humidity, are met.
5. Any side effects of the device or undesirable conditions resulting from its use shall be reasonably acceptable risks when weighed against the device’s intended performances.
6. In designing and constructing the device the manufacturer shall comply with principles of safety, taking account of the generally acknowledged state of the art at the time of manufacture.
7. The device shall be designed, manufactured and packed in a non—reusable packaging according to procedures which are sufficient to ensure that—
(a)the device is sterile when placed on the market; and
(b)if handled in accordance with conditions as to storage and transport laid down by the manufacturer, the device remains sterile until the packaging is removed and the device is implanted.
8. The device shall be designed and manufactured in such a way as to remove or minimise so far as is possible—
(a)the risk of physical injury arising out of its physical features (including its dimensions);
(b)risks connected with the use of sources of energy, including, in the case of a device which uses electricity, risks arising from lack of adequate insulation, from leakage currents and from overheating;
(c)risks connected with reasonably foreseeable environmental conditions such as magnetic fields, external electrical influences, electrostatic discharge, pressure, variations in pressure and acceleration;
(d)risks connected with medical treatment, in particular those resulting from the use of defibrillators or high—frequency surgical equipment;
(e)risks connected with ionising radiation from any radioactive substance included in the device in compliance with the protection requirements laid down in Directive 80/836/Eurotom(1) as amended by Directives 84/467/Euratom(2) and 84/466/Euratom(3);
(f)risks which may arise where maintenance and calibration are impossible, including—
(i)excessive increase of leakage currents,
(ii)ageing of the materials used,
(iii)excess heat generated by the device, and
(iv)decreased accuracy of any measuring or control mechanism.
9. The device shall be designed and manufactured in such a way as to guarantee the characteristics and performances referred to in paragraphs 1 to 5 with particular attention being paid to—
(a)the choice of materials used, particularly as regards toxicity;
(b)mutual compatibility between the materials used and biological tissues, cells and body fluids (acccount being taken in accordance with regulation 3(2) of the intended use of the device);
(c)compatibility of the device with any substances which it is intended to administer;
(d)the quality of the connections, particularly where considerations of safety are involved;
(e)the reliability of the source of energy;
(f)if appropriate, the need to make it leakproof;
(g)proper functioning of the programming and control systems, including software.
10. Where a device incorporates, as an integral part, a substance which, when used separately, is likely to be considered to be a medicinal product as defined in Article 1 of Directive 65/65/EEC(4), and whose action in combination with the device may result in its bioavailability, the safety, quality and usefulness of the substance, account being taken of the purpose of the device, must be verified by analogy with the appropriate methods specified in Directive 75/318/EEC(5), as last amended by Directive 91/507/EEC(6).
11. The device and, if appropriate, its component parts must be identifiable to allow any potential risk in connection with the device or any of its component parts.
12. The device shall bear a code by which it and its manufacturer can be unequivocably identified (particularly with regard to the type of device and year of manufacture) and which can be read, if necessary, without the need for a surgical operation.
13. When a device or its accessories bear instructions required for the operation of the device or indicate parameters for operating or adjusting it by means of a visual system, such instructions or indications must be understandable to the user and, as appropriate, the patient.
14.—(1) The sterile pack of the device must bear, legibly and indelibly, the following particulars where appropriate in the form of generally recognised symbols—
(a)the method of sterilisation;
(b)an indication that enables the sterile pack to be recognised as such;
(c)the name and address of the manufacturer;
(d)a description of the device;
(e)if the device is intended for clinical investigation, the words: “exclusively for clinical investigations”;
(f)if the device is custom—made, the words “custom—made device”;
(g)a declaration that the device is in a sterile condition;
(h)the month and year of manufacture;
(i)an indication of the time limit for implanting the device safely.
(2) The sales packaging of the device must bear, legibly and indelibly, the following particulars where appropriate in the form of generally recognised symbols—
(a)the name and address of the manufacturer;
(b)a description of the device;
(c)the purpose of the device;
(d)the relevant characteristics for its use;
(e)if the device is intended for clinical investigation, the words: “exclusively for clinical investigation”;
(f)if the device is custom—made, the words “custom—made device”;
(g)a declaration that the device is in a sterile condition;
(h)the month and year of manufacture;
(i)an indication of the time limit for implanting the device safely;
(j)the conditions for transporting and storing the device.
15.—(1) When placed on the market, each device must be accompanied by instructions for use giving the following particulars—
(a)the year of authorisation to affix the EC mark;
(b)the particulars specified in paragraph 14(1)(a) to (g) and (2)(c), (d) and (j);
(c)the performances referred to in paragraph 2 and any undesirable side effects;
(d)information allowing the registered medical practitioner to select a suitable device and the corresponding software and accessories;
(e)information constituting the instructions for use allowing the registered medical practitioner and, where appropriate, the patient to use the device, its accessories and software correctly;
(f)information on the nature of, and scope and times for, operating controls and trials;
(g)where appropriate, information on maintenance measures;
(h)information allowing, if appropriate, certain risks in connection with implantation of the device to be avoided;
(i)information regarding the risk of adverse effects on the device caused by instruments present at the time of specific investigations or treatment, and of adverse effects on instruments present at the time of those investigations or treatment caused by the device;
(j)the necessary instructions in the event of the sterile pack being damaged and, where appropriate, details of appropriate methods of resterilisation; and
(k)an indication, if appropriate, that a device can be reused only if it is reconditioned under the responsibility of the manufacturer to comply with the essential requirements.
(2) The instructions shall also include details allowing the registered medical practitioner to brief the patient on the contra—indications and the precautions to be taken, including, in particular—
(a)information allowing the lifetime of the energy source to be established;
(b)precautions to be taken should changes occur in the device’s performances;
(c)precautions to be taken as regards exposure, in reasonably foreseeable environmental conditions, to magnetic fields, external electrical influences, electrostatic discharge, pressure or variations in pressure, acceleration, and similar influences; and
(d)information identifying any medicinal products which the device in question is designed to administer.
Regulation 3(3)
1.—(1) Adequacy of the clinical data presented as part of the particulars referred to in paragraph 6(2) of Schedule 5 or included in the documentation mentioned in paragraphs 1(c) and 2 of Schedule 6 shall be based, account being taken as appropriate of relevant harmonised Standards, on either:
(a)a collation of currently available, relevant scientific literature covering the intended use of the device and the techniques for it as well as, if appropriate, a written report making a critical assessment of that collation; or
(b)the results of all clinical investigations made, and, in particular, the results of those carried out in accordance with paragraphs 2, 3 and 4.
(2) All the data presented must remain confidential unless it is essential that they be divulged.
2. The purposes of the clinical investigation are:
(a)to verify that, under normal conditions of use, the performances of the device comply with those indicated in paragraph 2 of Schedule 2; and
(b)to determine any undesirable side effects, under normal conditions of use, and assess whether they are reasonably acceptable risks having regard to the intended performance of the device.
3.—(1) Clinical investigations shall be made in accordance with the Declaration of Helsinki approved by the 18th World Medical Assembly in Helsinki, Finland, in 1964, and amended by the 29th World Medical Assembly in Tokyo, Japan, in 1975 and the 35th World Medical Assembly in Venice, Italy, in 1983.
(2) All measures relating to the protection of human subjects must be carried out in the spirit of the Declaration of Helsinki, as so approved and amended.
(3) Sub-paragraphs (1) and (2) apply to every step in the clinical investigation from first consideration of need and justification of the study to publication of results.
4.—(1) Clinical investigations shall—
(a)be performed according to an appropriate plan of investigation which represents the current state of the art and is defined in such a way as to confirm or refute the manufacturer’s claims for the device; and
(b)include an adequate number of observations to guarantee the scientific validity of the conclusions.
(2) The procedures used to perform the investigations shall be appropriate to the device under examination.
(3) Clinical investigations shall be performed in circumstances equivalent to those which would be found in normal conditions of use of the device.
(4) All appropriate features, including those involving the safety and performances of the device, and its effects on patients, shall be examined.
(5) All adverse events shall be fully recorded.
(6) The investigations shall be performed under the responsibity of an appropriately qualified medical specialist, in an appropriate environment.
(7) The medical specialist shall have access to the technical data regarding the device.
(8) The written report, signed by the responsible medical specialist, shall comprise a critical evaluation of all the data collected during the clinical investigation.
Regulations 3(7) and (8), 6(2) and 7(1)
1.—(1) The manufacturer or his authorised representative shall draw up for custom-made devices or for devices intended for clinical investigation a written statement in English comprising the particulars specified in sub-paragraph (1) or (2) respectively of paragraph 2.
(2) The statement shall be signed by or on behalf of the manufacturer or by his authorised representative.
2.—(1) The particulars referred to in paragraph 1(1) for a custom-made device are—
(a)particulars allowing the device to be identified;
(b)a statement affirming that the device is intended for exclusive use by a particular patient, together with his name;
(c)the name of the registered medical practitioner who wrote the prescription and, if applicable, the name of the clinic concerned;
(d)the particular features of the device as described in the prescription;
(e)a statement affirming that the device complies with the essential requirements and, where applicable, indicating which relevant essential requirements have not been wholly complied with, together with the grounds for them not having been wholly complied with.
(2) The particulars referred to in paragraph 1(1) for a device intended for clinical investigation are—
(a)particulars allowing the device to be identified;
(b)a plan of the investigation, stating in particular the purpose, scope and number of the devices concerned, the terms of any opinion on the investigation given by a committee or other body which has considered the ethics of the investigation and a description of the precautions taken to protect the health and safety of the patient;
(c)the name of each registered medical practitioner and of each institution responsible for the investigation;
(d)each place, date of commencement and duration scheduled for the investigation;
(e)a statement affirming that the device complies with the relevant essential requirements apart from the aspects constituting the object of the investigations and that, with regard to those aspects, every precaution has been taken to protect the health and safety of the patient.
3. The manufacturer shall undertake to keep available for the Secretary of State—
(a)for custom-made devices, documentation in English enabling the design, manufacture and performances of the product, including the expected performances, to be understood, so as to allow conformity with the requirements of the Directive to be assessed;
(b)for devices intended for clinical investigation, that documentation and the following additional documentation in English:—
(i)a general description of the product,
(ii)design drawings, particulars of methods of manufacture, in particular as regards sterilisation, and diagrams of parts, sub-assemblies, circuits and similar items,
(iii)the descriptions and explanations necessary for the understanding of those drawings and diagrams and of the operation of the product,
(iv)a list of the relevant national Standards applied in full or in part, and a description of the measures taken to comply with the relevant essential requirements where those Standards have not been applied,
(v)the results of the design calculations, checks and technical tests carried out.
Regulation 5(2)
1. A manufacturer who wishes to have his quality system for design, manufacture and final inspection approved shall make an application in writing to a notified body for evaluation of his quality system.
2.—(1) The application shall include—
(a)a description of the type of devices manufacture of which is envisaged;
(b)details of any relevant national Standard or harmonised Standard with which the devices comply;
(c)the quality system documentation;
(d)an undertaking by the manufacturer to fulfil the obligations arising from the quality system as approved;
(e)an undertaking by the manufacturer to maintain the approved quality system in such a way that it remains adequate and efficacious;
(f)an undertaking by the manufacturer to institute and keep up-dated a post-marketing surveillance system.
(2) The undertaking referred to in sub-paragraph (1)(f) shall include an obligation for the manufacturer to notify the Secretary of State of the following incidents immediately on learning of them—
(a)any deterioration in the characteristics or performances of, and any inaccuracies in the instruction leaflet for, a device which might lead to or have led to the death of a patient or a deterioration in the state of his health;
(b)any technical or medical reason for withdrawal of a device from the market which results in such withdrawal by the manufacturer.
3.—(1) Approval of a quality system shall not be given unless the conditions specified in subparagraphs (2) to (4) are fulfilled.
(2) The application of the quality system must ensure that the devices conform to the provisions of the Directive which apply to them at every stage, from design to final controls.
(3) All the elements, requirements and provisions adopted by the manufacturer for his quality system shall be documented in a systematic and orderly manner in the form of written policies and procedures, and the documentation must make possible a uniform interpretation of the policies and procedures as to quality such as those programmes, plans, manuals and records which relate to quality.
(4) The documentation referred to in sub-paragraph (3) shall include in particular an adequate description of—
(a)the manufacturer’s objectives as to quality;
(b)the organisation of the business and in particular—
(i)the organisational structure of the organisation and the responsibilities and degrees of authority of the managerial staff, where quality of design and manufacture of devices is concerned;
(ii)the methods of monitoring the efficient operation of the quality system and in particular its ability to achieve the desired quality of the design and of the devices, including control of devices which do not conform to harmonised Standards;
(c)the procedures for monitoring and verifying the design of the products and in particular—
(i)the design specifications, including the standards which will be applied and a description of the solutions adopted to fulfil the relevant essential requirements when the relevant national Standards are not applied in full;
(ii)the techniques of control and verification of the design, and the processes and systematic actions which will be used when the devices are being designed;
(d)the techniques of control and of quality assurance at the manufacturing stage and in particular:
(i)the processes and procedures which will be used, particularly as regards sterilization, purchasing and the relevant documents,
(ii)product-identification procedures drawn up and kept up to date from drawings, specifications or other relevant documents at every stage of manufacture;
(e)the appropriate tests and trials which will be effected before, during and after production, the frequency with which they will take place, and the equipment used for testing.
4.—(1) The notified body shall effect an evaluation of the quality system to determine whether it fulfils the conditions specified in paragraph 3(2) to (4), and in doing so shall presume fulfilment of those conditions for a quality system which is in accordance with relevant harmonized Standards.
(2) The team of individuals entrusted with the evaluation shall include at least one person who has already had experience of evaluations of the technology concerned.
(3) The evaluation shall include an inspection at the manufacturer’s premises.
(4) A decision as to whether or not approval is given shall be notified to the manufacturer after the final inspection and shall contain a reasoned evaluation and the conclusions of the notified body.
5.—(1) The manufacturer shall inform the relevant notified body of any plan to modify his approved quality system.
(2) The relevant notified body shall evaluate the proposed modifications and shall determine whether the quality system as proposed to be modified would fulfil the conditions specified in paragraph 3(2) to (4).
(3) A decision as to whether or not to approve the system as modified shall be notified to the manufacturer and shall contain a reasoned evaluation and the conclusions of the relevant notified body.
(4) In this paragraph “relevant notified body” in relation to a quality system means the notified body which approved it, except that where that body is no longer a notified body as respects tasks under this paragraph, it means such other notified body as the Secretary of State shall have designated as the notified body in relation to those tasks and that system.
6.—(1) A manufacturer who makes an application under paragraph 1 relating to a type of device shall also make an application for examination of the design dossier relating to that type of device which he plans to manufacture.
(2) The application shall describe the design, manufacture, and performances of the device in question and shall include the necessary particulars which make it possible to evaluate whether it complies with the requirements of the Directive.
(3) The particulars referred to in sub-paragraph (2) shall include the following:—
(a)the design specifications, including the Standards which have been applied;
(b)the necessary proof of their appropriateness, in particular where the relevant national Standards have not been applied in full, including the results of the appropriate tests carried out by the manufacturer or under his responsibility;
(c)a statement as to whether or not the device incorporates, as an integral part, a substance referred to in paragraph 10 of Schedule 2, whose action in combination with the device may result in its bioavailability, together with data on the relevant tests conducted;
(d)the clinical data referred to in Schedule 3;
(e)the draft instruction leaflet.
7.—(1) The notified body shall examine the application and, where the type of device complies with the relevant provisions of the Directive, shall signify approval by issuing the applicant with an EC design examination certificate.
(2) The notified body may require the application to be supplemented by further tests or proof so that compliance with the requirements of the Directive may be evaluated.
(3) The certificate shall contain the conclusions of the examination, the conditions of its validity, the data needed for identification of the approved design and, where appropriate, a description of the intended use of the type of device.
8.—(1) An applicant to whom an EC design examination certificate has been issued shall inform the relevant notified body of any modification made to the approved design.
(2) Any such modification must receive approval from the relevant notified body where it may affect conformity with the relevant essential requirements or the conditions prescribed for the use of the type of device.
(3) Any supplementary approval which is given shall be given in the form of an addendum to the EC design examination certificate.
(4) In this paragraph and paragraph 9 “relevant notified body” in relation to a design means the notified body which issued the EC design examination certificate in respect of it, except that where that body is no longer a notified body as respects tasks under this paragraph it means such other notified body as the Secretary of State shall have designated as the notified body in relation to those tasks and that design.
9. The relevant notified body—
(a)shall withdraw any EC design examination certificate or supplementary approval given by it if it considers that the device no longer conforms with the relevant essential requirements or the conditions prescribed for the use of the type of device; and
(b)may withdraw the EC design examination certificate or supplementary approval if it has been given on the basis of false or misleading information.
10.—(1) The manufacturer whose quality system for manufacture and inspection is approved shall authorise the notified body to carry out the inspections, and shall supply it with all the appropriate information, which the notified body needs to ensure that the manufacturer duly fulfils his obligations arising from the approved quality system, and that information shall include—
(a)the documentation relating to the quality system;
(b)the data stipulated in the part of the quality system relating to design, such as the results of analyses, calculations, tests, etc.;
(c)the data stipulated in the part of the quality system relating to manufacture, such as reports concerning inspections, tests, standardisations, calibrations, the qualifications of the staff concerned, etc.
(2) The notified body shall periodically carry out appropriate evaluations, including inspections, in order to ascertain that the manufacturer is applying the approved quality system, and shall supply the manufacturer with a report about any such evaluation.
(3) In addition, the notified body may make inspections on unannounced visits to the manufacturer, and shall supply him with a report about any such inspection.
(4) Every notified body shall communicate to the other notified bodies the information necessary to establish in respect of which quality systems it has issued, refused or withdrawn approval.
Regulation 5(3)
1.—(1) An application for EC type-examination shall be made in writing by the manufacturer, or by his authorised representative, to a notified body.
(2) The application shall include—
(a)the name and address of the manufacturer and the name and address of the authorised representative if the application is made by the latter;
(b)a written declaration that an application has not been made to any other notified body;
(c)the documentation specified in paragraph 2 needed to allow an evaluation to be made of the conformity of a representative sample of the production in question (hereinafter referred to as “type”), with the requirements of the Directive.
(3) The applicant shall make available to the notified body a type and such other samples as the notified body may request.
2. The documentation referred to in paragraph 1(2)(c) is that which makes it possible to understand the design, the manufacture and the performances of the device, containing in particular the following items:
(a)a general description of the device;
(b)design drawings, a description of the methods of manufacture envisaged, in particular as regards sterilisation, and diagrams of parts, sub-assemblies, circuits and other compo nents and accessories;
(c)the descriptions and explanations necessary for the understanding of those drawings and diagrams and of the operation of the device;
(d)a list of the relevant national Standards applied in full or in part, and a description of the solutions adopted to satisfy the relevant essential requirements where the relevant national Standards have not been applied;
(e)the results of design calculations, investigations and technical tests carried out and other design verification data;
(f)a statement as to whether or not the device incorporates, as an integral part, a substance described in paragraph 10 of Schedule 2 whose action in combination with the device may result in its bioavailability, together with data on the relevant trials conducted;
(g)clinical data in accordance with paragraph 1 of Schedule 3;
(h)the draft instruction leaflet.
3. The notified body shall—
(a)examine and evaluate the documentation;
(b)verify that the type has been manufactured in accordance with that documentation;
(c)record the items which have been designed in accordance with the applicable provisions of the relevant national Standards, and the items which have not been so designed;
(d)where the relevant national Standards have not been applied, carry out or have carried out the inspections and the tests necessary to verify whether the solutions adopted by the manufacturer satisfy the relevant essential requirements;
(e)where the manufacturer has chosen to apply the relevant national Standards, carry out or have carried out the inspections and the tests necessary to verify whether they have actually been applied;
(f)seek agreement with the applicant on the place where the necessary inspections and tests will be carried out.
4.—(1) Where the type complies with the provisions of the Directive, the notified body shall signify approval by issuing an EC type-examination certificate to the applicant.
(2) A certificate relating to a type shall contain the name and address of the manufacturer, the conclusions of the evaluation of the type, the conditions under which the certificate is valid and the information necessary for identification of the type.
(3) The notified body shall attach to the certificate what it considers to be the significant parts of the documentation and shall keep a copy of the certificate and of those parts of the documentation.
5.—(1) The applicant shall inform the relevant notified body of any modification made to a device in relation to which an EC type-examination certificate has been issued.
(2) Any such modification must receive approval from the relevant notified body where it may affect conformity with the relevant essential requirements or with the conditions of use specified for the device.
(3) Any such approval shall be issued, where appropriate, in the form of a supplement to the initial EC type-examination certificate.
(4) In this paragraph and paragraph 6 “relevant notified body” in relation to a device means the notified body which issued the EC type-examination certificate in respect of it, except that where that body is no longer a notified body as respects tasks under this paragraph it means such other notified body as the Secretary of State shall have designated in relation to that device or devices of that type.
6. The relevant notified body—
(a)shall withdraw any EC type-examination certificate or supplementary approval given by it if it considers that the device no longer conforms with the relevant essential requirements or the conditions of use specified for the device; and
(b)may withdraw the EC type-examination certificate or supplementary approval if it has been given on the basis of false or misleading information.
7. Each notified body shall communicate to the other notified bodies the information necessary to establish in respect of which devices it has issued, refused or withdrawn EC type-examination certificates and supplements.
8.—(1) A notified body shall, if so requested by another notified body, provide that other body with a copy of EC type-examination certificates, and of any supplements to them, which it has issued.
(2) A notified body shall make available to another notified body which so requests, giving reasons for its request, copies of those parts of the documentation which are attached to a certificate under paragraph 4(3), but shall not do so without having first informed the manufacturer.
Regulation 5(4)
1. The manufacturer shall, before the start of manufacture, prepare and submit to the notified body documentation defining the manufacturing process, in particular as regards sterilisation, and all the routine, pre-established procedures to be implemented to ensure homogeneity of production and conformity of the devices with—
(a)the type described in the EC type-examination certificate; and
(b)the relevant requirements of the Directive.
2.—(1) The manufacturer shall undertake to the notified body to institute and keep up-dated a system of post-marketing surveillance.
(2) The undertaking shall include an obligation for the manufacturer to notify the Secretary of State of the following events immediately on learning of them:
(a)any deterioration in the characteristics or performances of and any inaccuracies in the instruction leaflet for, a device which might lead to or have led to the death of a patient or a deterioration in his state of health;
(b)any technical or medical reason resulting in withdrawal of a device from the market by the manufacturer.
3.—(1) The notified body shall carry out EC verification by controls and tests on the products on a statistical basis in accordance with paragraph 4.
(2) The manufacturer shall authorise the notified body to evaluate the efficiency of the measures taken pursuant to paragraph 1, by audit where appropriate.
4.—(1) The manufacturer shall present the manufactured products to the notified body in the form of homogeneous batches.
(2) A random sample shall be taken by the notified body from each batch.
(3) The devices which make up the sample shall be examined individually and appropriate tests, defined in the relevant national Standards or equivalent tests, shall be carried out to verify the conformity of the devices with the type described in the EC type-examination certificate, in order to determine whether the batch is to be accepted or rejected.
(4) Statistical control of products will be based on attributes, entailing a sampling system with the following characteristics:
(a)a level of quality corresponding to a probability of acceptance of 95%, with a non-conformity percentage of between 0.29 and 1%;
(b)a limit quality corresponding to a probability of acceptance of 5%, with a non-conformity percentage of between 3 and 7%.
(5) If a batch is accepted, the notified body shall draw up a written certificate of conformity, and all the devices in the batch may, subject to regulation 3(1), be placed on the market, with the exception of those devices in the sample which were found not to conform.
(6) If a batch is rejected, the notified body shall take the appropriate measures to prevent the batch from being placed on the market.
(7) If justified on practical grounds, the manufacturer may affix the EC mark during manufacture, under the responsibility of the notified body, in accordance with regulation 4(1), accompanied by the identifying logo of the notified body responsible for statistical verification.
Regulation 5(5)
1.—(1) The manufacturer shall make an application in writing for evaluation of his quality system to a notified body.
(2) The application shall be signed by or on behalf of the manufacturer and shall include—
(a)the information and undertakings which are required by paragraph 2 of Schedule 5 to be included in an application under paragraph 1 of that Schedule, other than the information required by paragraph 2(1)(e) of that Schedule so far as it relates to research and development and design;
(b)the technical documentation relating to the approved type and a copy of the EC type-examination certificate.
2.—(1) Application of the quality system must ensure that the products conform to the type described in the EC type-examination certificate.
(2) Subject to sub-paragraph (3) below, paragraphs 3(3) and (4)(a), (b), (d) and (e), 4 and 5 of Schedule 5 shall apply for the purposes of this Schedule as they apply for the purpoes of that Schedule, but as though any reference to paragraph 3(2) to (4) of that Schedule were a reference to sub-paragraph (1) of this paragraph and paragraphs 3(3) and (4)(a), (b), (d) and (e) of that Schedule.
(3) Paragraph 3(4)(b)(ii) of Schedule 5 shall not apply in respect of the desired quality of design.
3. The aim of surveillance is to ensure that the manufacturer duly fulfils the obligations which arise from the approved quality system.
4. The manufacturer shall authorise the notified body to carry out all necessary inspections and shall supply it with all appropriate information, in particular:
(a)the documentation relating to the quality system documentation;
(b)the data stipulated in the part of the quality system relating to manufacture, such as reports concerning inspections, tests, standardisations, calibrations, the qualifications of the staff concerned, etc.
5.—(1) The notified body shall periodically carry out appropriate inspections and evaluations in order to ascertain that the manufacturer is applying the approved quality system, and shall supply the manufacturer with an evaluation report.
(2) In addition, the notified body may make unannounced visits to the manufacturer, and shall supply him with an inspection report.
(3) The notified body shall communicate to the other notified bodies all relevant information concerning approvals of quality systems issued, refused or withdrawn.
Regulation 8
1.—(1) The body, its director and its staff responsible for carrying out operations relating to evaluation and verification—
(a)shall neither be, nor be the authorised representative of, the designer, manufacturer, supplier or installer of devices which they evaluate;
(b)shall neither be, nor represent any person who is, directly involved in the design, construction, marketing or maintenance of those devices.
(2) Nothing in sub-paragraph (1) precludes the possibility of exchanges of technical information between a manufacturer and the body.
2. The body and its staff must—
(a)carry out operations relating to evaluation and verification with the highest degree of professional integrity and technical competence;
(b)be free from all pressures and inducements, particularly financial, which might influence their judgment or the results of an evaluation, especially from those with an interest in the result of the evalution.
3.—(1) The body must be able to carry out all the tasks as respects which it is or is to be designated under regulation 8, whether they are or are to be carried out by the body itself or under its responsibility.
(2) In particular the body must—
(a)have at its disposal the necessary staff; and
(b)possess the necessary facilities,
to enable it to perform properly the technical and adminstrative tasks relating to evaluation and verification; and must have access to the equipment necessary for the verifications required.
4. The staff responsible for evaluation and verification must have:
(a)appropriate vocational training covering all the operations of evaluation and verification for which the body has been designated;
(b)satisfactory knowledge of the requirements of the controls they carry out and adequate experience of the operations of evaluation and verification for which the body has been designated;
(c)the ability required to draw up the certificates, records and reports to demonstrate that the controls have been carried out.
5. The impartiality of staff carrying out evaluations and verifications must be guaranteed, and their remuneration must not depend on the number of procedures carried out, nor on the results of the evaluations and verifications.
6. The body must keep itself insured against liabilities incurred in the performance of its functions, except to the extent that any liability would be met by the Crown, a government Department or a body exercising functions on behalf of the Crown.
7. The staff of the body must observe professional secrecy with regard to all information gained in carrying out their tasks (except vis-a-vis the competent authorities of any member State in which their activities are carried out) under the Directive or these Regulations.
8. The body must undertake to carry out the inspections referred to in paragraph 10(2) of Schedule 5 with a frequency to be agreed with the Secretary of State, and, once having become a notified body must fulfil the terms of that undertaking.
9. The body must undertake to comply with Articles 13 and 15 of the Directive (measures where EC mark has been wrongly affixed and duty of confidentiality), and, once having become a notified body, must fulfil the terms of that undertaking.
OJ No. L246, 17.9.80, p. 1.
OJ No. L265, 5.10.84, p. 4.
OJ No. L265, 5.10.84, p. 1.
OJ No. 22, 9.2.65, p. 369.
OJ No. L147, 9.6.75, p. 1.
OJ No. L270, 26.9.91, p. 32
Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.
Original (As Enacted or Made): The original version of the legislation as it stood when it was enacted or made. No changes have been applied to the text.
Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include: