[F1SCHEDULE 1GE+W+SAUTOMATIC WEIGHING INSTRUMENTS (MI-006) (Annex VIII to the Directive)

The relevant essential requirements of Schedule 1A, the specific requirements of this Schedule and the conformity assessment procedures listed in Chapter I of this Schedule, apply to automatic weighing instruments defined below, intended to determine the mass of a body by using the action of gravity on that body.

DEFINITIONS

Automatic weighing instrumentAn instrument that determines the mass of a product without the intervention of an operator and follows a predetermined programme of automatic processes characteristic of the instrument.
Automatic catchweigherAn automatic weighing instrument that determines the mass of pre-assembled discrete loads (for example prepackages) or single loads of loose material.
Weight labellerAn automatic catchweigher that labels individual articles with the weight value.
Weight/price labellerAn automatic catchweigher that labels individual articles with the weight value, and price information.
Automatic gravimetric filling instrumentAn automatic weighing instrument that fills containers with a predetermined and virtually constant mass of product from bulk.
Discontinuous totaliser (totalising hopper weigher)An automatic weighing instrument that determines the mass of a bulk product by dividing it into discrete loads. The mass of each discrete load is determined in sequence and summed. Each discrete load is then delivered to bulk.
Continuous totaliserAn automatic weighing instrument that continuously determines the mass of a bulk product on a conveyor belt, without systematic subdivision of the product and without interrupting the movement of the conveyor belt.
Rail-weighbridgeAn automatic weighing instrument having a load receptor inclusive of rails for conveying railway vehicles.

SPECIFIC REQUIREMENTS

CHAPTER IE+W+SRequirements common to all types of automatic weighing instruments

Rated Operating ConditionsE+W+S

1.  The manufacturer shall specify the rated operating conditions for the instrument as follows:

1.1.  For the measurand:E+W+S

The measuring range in terms of its maximum and minimum capacity.

1.2.  For the electrical supply influence quantities:E+W+S

In case of AC voltage supply:the nominal AC voltage supply, or the AC voltage limits.
In case of DC voltage supply:the nominal and minimum DC voltage supply, or the DC voltage limits.

1.3.  For the mechanical and climatic influence quantities:E+W+S

The minimum temperature range is 30 °C unless specified otherwise in the following chapters of this Schedule.

The mechanical environment classes according to Schedule 1A, paragraph 1.3.2 are not applicable. For instruments which are used under special mechanical strain, e.g. instruments incorporated into vehicles, the manufacturer shall define the mechanical conditions of use.

1.4  For other influence quantities (if applicable):E+W+S

2.  Permissible effect of disturbances — Electromagnetic environmentE+W+S

The required performance and the critical change value are given in the relevant Chapter of this Schedule for each type of instrument.

SuitabilityE+W+S

3.1.  Means shall be provided to limit the effects of tilt, loading and rate of operation such that maximum permissible errors (MPEs) are not exceeded in normal operation.E+W+S

3.2.  Adequate material handling facilities shall be provided to enable the instrument to respect the MPEs during normal operation.E+W+S

3.3.  Any operator control interface shall be clear and effective.E+W+S

3.4.  The integrity of the display (where present) shall be verifiable by the operator.E+W+S

3.5.  Adequate zero setting capability shall be provided to enable the instrument to respect the MPEs during normal operation.E+W+S

3.6.  Any result outside the measurement range shall be identified as such, where a printout is possible.E+W+S

Conformity assessmentE+W+S

4.  The conformity assessment procedures specified in the modules in Schedule 1B applicable to automatic weighing instruments that the manufacturer can choose between are:

(a)For mechanical systems:

(i)B and D;

(ii)B and E;

(iii)B and F;

(iv)D1;

(v)F1;

(vi)G; or

(vii)H1.

(b)For electromechanical instruments:

(i)B and D;

(ii)B and E;

(iii)B and F;

(iv)G; or

(v)H1.

(c)For electronic systems or systems containing software:

(i)B and D;

(ii)B and F;

(iii)G; or

(iv)H1.

CHAPTER IIE+W+SAutomatic Catchweighers

1.  These categories are divided into four accuracy classes:E+W+S

Y(I), Y(II), Y(a) & Y(b)

which shall be specified by the manufacturer.

MPEE+W+S

2.1.  MPE Category Y instrumentsE+W+S

Table 1

Net Load (m) in verification scale intervals (e)Maximum permissible mean errorMaximum permissible error
Y(I)Y(II)Y(a)Y(b)StaticAutomatic
0 < m ≤ 50,0000 < m ≤ 5,0000 < m ≤ 5000 < m ≤ 50± 0.5 e± 1 e
50,000 < m ≤ 200,0005,000 < m ≤ 20,000500 < m ≤ 2,00050 < m ≤ 200± 1.0 e± 1.5 e
200,000 < m20,000 < m ≤ 100,0002,000 < m ≤ 10,000200 < m ≤ 1,000± 1.5± 2 e

Verification scale interval — single interval instrumentsE+W+S

Table 2
Accuracy classesVerification scale intervalNumber of verification scale intervalsn = Max/e
MinimumMaximum
MinimumMaximum
XIY(I)0.001 g ≤ e50,000
XIIY(II)0.001 g ≤ e ≤ 0.05 g100100,000
0.1 g ≤ e5,000100,000
XIIIY(a)0.1 g ≤ e ≤ 2 g10010,000
5 g ≤ e50010,000
XIIIIY(b)5 g ≤ e1001,000

Verification scale interval — multi-interval instrumentsE+W+S

Table 3

1 For i = r the corresponding column of Table 2 applies with e replaced by er.

Verification scale intervalNumber of verification scale intervalsn = Max/e

Minimum value1

n = Maxi /e(i+1)

For i = r the corresponding column of Table 2 applies with e replaced by er.

Maximum value

n = Maxi /ei

Y(I)0.001 g ei50,000
Y(II)0.001 g ei 0.05 g5,000100,000
0.1 g ei5,000100,000
Y(a)0.1 g ei50010,000
Y(b)5 g ei501 000

Where:

Measurement RangeE+W+S

3.  In specifying the measurement range for class Y instruments the manufacturer shall take account that the minimum capacity shall not be less than:

class Y(I):100 e
class Y(II):20 e for 0.001 g ≤ e ≤ 0.05 g, and 50 e for 0.1 g ≤ e
class Y(a):20 e
class Y(b):10 e
Scales used for grading, e.g. postal scales and garbage weighers:5 e

Dynamic SettingE+W+S

4.1.  The dynamic setting facility shall operate within a load range specified by the manufacturer.E+W+S

4.2.  When fitted, a dynamic setting facility that compensates for the dynamic effects of the load in motion shall be inhibited from operating outside the load range, and shall be capable of being secured.E+W+S

Performance Under Influence Factors And Electromagnetic DisturbancesE+W+S

5.1.  The MPEs due to influence factors are:E+W+S

5.1.1.  For category Y instrumentsE+W+S

5.1.2.  The critical change value due to a disturbance is one verification scale interval.E+W+S

5.2.  Temperature range:E+W+S

CHAPTER IIIE+W+SAutomatic Gravimetric Filling Instruments

Accuracy classesE+W+S

1.1.  The manufacturer shall specify both the reference accuracy class Ref(x) and the operational accuracy class(es) X(x).E+W+S

1.2.  An instrument type is designated a reference accuracy class, Ref(x), corresponding to the best possible accuracy for instruments of the type. After installation, individual instruments are designated for one or more operational accuracy classes, X(x), having taken account of the specific products to be weighed. The class designation factor (x) shall be ≤ 2, and in the form 1 × 10k, 2 × 10k or 5 × 10k where k is a negative whole number or zero.E+W+S

1.3.  The reference accuracy class, Ref(x) is applicable for static loads.E+W+S

1.4.  For the operational accuracy class X(x), X is a regime relating accuracy to load weight and (x) is a multiplier for the limits of error specified for class X(1) in paragraph 2.2.E+W+S

MPEE+W+S

Static weighing errorE+W+S

2.1.1.  For static loads under rated operating conditions, the MPE for reference accuracy class Ref(x), shall be 0.312 of the maximum permissible deviation of each fill from the average; as specified in Table 5; multiplied by the class designation factor (x).E+W+S

2.1.2.  For instruments where the fill may be made up from more than one load (e.g. cumulative or selective combination weighers) the MPE for static loads shall be the accuracy required for the fill as specified in paragraph 2.2 (i.e. not the sum of the maximum permissible deviation for the individual loads).E+W+S

Deviation from average fillE+W+S

Table 4
Value of the mass, m (g), of the fillsMaximum permissible deviation of each fill from the average for class X(1)
m ≤ 507.2 %
50 < m ≤ 1003.6 g
100 < m ≤ 2003.6 %
200 < m ≤ 3007.2 g
300 < m ≤ 5002.4 %
500 < m ≤ 1,00012 g
1,000 < m ≤ 10,0001.2 %
10,000 < m ≤ 15,000120 g
15,000 < m0.8 %

Note:

The calculated deviation of each fill from the average may be adjusted to take account for the effect of material particle size.

Error relative to pre-set value (setting error)E+W+S

2.3.  For instruments where it is possible to pre-set a fill weight; the maximum difference between the pre-set value and the average mass of the fills shall not exceed 0.312 of the maximum permissible deviation of each fill from the average, as specified in Table 4.

Performance Under Influence Factor And Electromagnetic DisturbanceE+W+S

3.1.  The MPE due to influence factors shall be as specified in paragraph 2.1.E+W+S

3.2.  The critical change value due to a disturbance is a change of the static weight indication equal to the MPE as specified in paragraph 2.1 calculated for the rated minimum fill, or a change that would give equivalent effect on the fill in the case of instruments where the fill consists of multiple loads. The calculated critical change value shall be rounded to the next higher scale interval (d).E+W+S

3.3.  The manufacturer shall specify the value of the rated minimum fill.E+W+S

CHAPTER IVE+W+SDiscontinuous Totalisers

1.  Accuracy ClassesE+W+S

Instruments are divided into four accuracy classes as follows: 0.2; 0.5; 1; 2.

2.  MPEsE+W+S

Table 5

Accuracy classMPE of totalised load
0.2± 0.10 %
0.5± 0.25 %
1± 0.50 %
2± 1.00 %

Totalisation scale intervalE+W+S

3.  The totalisation scale interval (dt) shall be in the range: 0.01 % Max ≤ dt ≤ 0.2 % Max

Minimum Totalised Load (Σmin)E+W+S

4.  The minimum totalised load (Σmin) shall be not less than the load at which the MPE is equal to the totalisation scale interval (dt) and not less than the minimum load as specified by the manufacturer.

Zero SettingE+W+S

5.  Instruments that do not tare weigh after each discharge shall have a zero setting device. Automatic operation shall be inhibited if zero indication varies by:

Operator InterfaceE+W+S

6.  Operator adjustments and reset function shall be inhibited during automatic operation.

PrintoutE+W+S

7.  On instruments equipped with a printing device, the reset of the total shall be inhibited until the total is printed. The printout of the total shall occur if automatic operation is interrupted.

Performance under influence factors and electromagnetic disturbancesE+W+S

8.1.  The MPEs due to influence factors shall be as specified in Table 6.E+W+S

Table 6

Load (m) in totalisation scale intervals (dt)MPE
0 < m ≤ 500± 0.5 dt
500 < m ≤ 2,000± 1.0 dt
2,000 < m ≤ 10,000± 1.5 dt

8.2.  The critical change value due to a disturbance is one totalisation scale interval for any weight indication and any stored total.E+W+S

CHAPTER VE+W+S

Accuracy classesE+W+S

1.  Instruments are divided into three accuracy classes as follows: 0.5; 1; 2.

Measurement RangeE+W+S

2.1.  The manufacturer shall specify the measurement range, the ratio between the minimum net load on the weighing unit and the maximum capacity, and the minimum totalised load.E+W+S

2.2.  The minimum totalised load Σmin shall not be less thanE+W+S

Where d is the totalisation scale interval of the general totalisation device.

MPEE+W+S

Table 7
Accuracy classMPE for totalised load
0.5± 0.25 %
1± 0.5 %
2± 1.0 %

Speed of the beltE+W+S

4.  The speed of the belt shall be specified by the manufacturer. For single-speed beltweighers, and variable-speed beltweighers having a manual speed setting control, the speed shall not vary by more than 5 % of the nominal value. The product shall not have a different speed than the speed of the belt.

General Totalisation DeviceE+W+S

5.  It shall not be possible to reset the general totalisation device to zero.

Performance under influence factors and electromagnetic disturbancesE+W+S

6.1.  The MPE due to influence factor, for a load not less than the Σmin, shall be 0.7 times the appropriate value specified in Table 7, rounded to the nearest totalisation scale interval (d).E+W+S

6.2.  The critical change value due to a disturbance shall be 0.7 times the appropriate value specified in Table 7, for a load equal to Σmin, for the designated class of the beltweigher; rounded up to the next higher totalisation scale interval (d).E+W+S

CHAPTER VIE+W+SAutomatic Rail Weighbridges

Accuracy classesE+W+S

1.  Instruments are divided into four accuracy classes as follows: 0.2; 0.5; 1; 2.

MPEE+W+S

2.1.  The MPEs for weighing-in-motion of a single wagon or a total train are shown in Table 8.E+W+S

Table 8

Accuracy classMPE
0.2± 0.1 %
0.5± 0.25 %
1± 0.5 %
2± 1.0 %

2.2.  The MPEs for the weight of coupled or uncoupled wagons weighing-in-motion shall be one of the following values, whichever is the greatest:E+W+S

2.3.  The MPEs for the weight of train weighing-in-motion shall be one of the following values, whichever is the greatest:E+W+S

2.4.  When weighing coupled wagons; the errors of not more than 10 % of the weighing results taken from one or more passes of the train may exceed the appropriate MPE given in paragraph 2.2, but shall not exceed twice the MPE.E+W+S

Scale interval (d)E+W+S

3.  The relationship between the accuracy class and the scale interval shall be as specified in Table 9.

Table 9
Accuracy classScale interval (d)
0.2d ≤ 50 kg
0.5d ≤ 100 kg
1d ≤ 200 kg
2d ≤ 500 kg

Measurement rangeE+W+S

4.1.  The minimum capacity shall not be less than 1 t, and not greater than the value of the result of the minimum wagon weight divided by the number of partial weighings.E+W+S

4.2.  The minimum wagon weight shall not be less than 50 d.E+W+S

Performance under influence factor and electromagnetic disturbanceE+W+S

5.1.  The MPE due to an influence factor shall be as specified in Table 10.E+W+S

Table 10

Load (m) in verification scale intervals (d)MPE
0 < m ≤ 500± 0.5 d
500 < m ≤ 2,000± 1.0 d
2,000 < m ≤ 10,000± 1.5 d

5.2.  The critical change value due to a disturbance is one scale interval.]E+W+S