Rated operating conditionsU.K.
14. The manufacturer shall specify the rated operating conditions of the relevant instrument, in particular—
(a)the values of fn, Un, In, Ist, Imin, Itr and Imax that apply to the relevant instrument. For the current values specified, the relevant instrument shall satisfy the conditions given in Table 2;
Table 2
1 For Class B electromechanical relevant instruments Imin ≤ 0.4 · Itr shall apply. | |||
Class A | Class B | Class C | |
---|---|---|---|
For direct-connected relevant instruments | |||
Ist | ≤ 0.05 · Itr | ≤ 0.04 · Itr | ≤ 0.04 · Itr |
Imin | ≤ 0.5 · Itr | ≤ 0.5 · Itr | ≤ 0.3 · Itr |
Imax | ≥ 50 · Itr | ≥ 50 · Itr | ≥ 50 · Itr |
For transformer-operated relevant instruments | |||
Ist | ≤ 0.06 · Itr | ≤ 0.04 · Itr | ≤ 0.02 · Itr |
Imin | ≤ 0.4 · Itr | ≤ 0.2 · Itr1 | ≤ 0.2 · Itr |
In | = 20 · Itr | = 20 · Itr | = 20 · Itr |
Imax | ≥ 1.2 · In | ≥ 1.2 · In | ≥ 1.2 · In |
(b)the voltage, frequency and power factor ranges within which the relevant instrument shall satisfy the MPE requirements are specified in Table 3. These ranges shall recognise the typical characteristics of electricity supplied by public distribution systems;
(c)the voltage and frequency ranges shall be at least—
(i)0.9 · Un ≤ U ≤ 1.1 · Un;
(ii)0.98 · fn ≤ f ≤ 1.02 · fn;
(d)the power factor range shall be at least from cosφ = 0.5 inductive to cosφ = 0.8 capacitive.