1. The units of mass used shall be the legal units within the meaning of Directive 80/181/EEC(1) as last amended by Directive 1999/103/EC(2).
Subject to compliance with this condition, the following units are permitted:
SI units: kilogram, microgram, milligram, gram, tonne,
Imperial units: Troy ounce,
other non-SI units: metric carat, if weighing precious stones.
For instruments that make use of the Imperial units of mass referred to above, the relevant essential requirements specified below shall be converted to the said Imperial units, using simple interpolation.
2.1. I special
II high
III medium
IIII ordinary
The specifications of these classes are given in Table 1.
The minimum capacity is reduced to 5e for instruments in classes II and III for determining a conveying tariff.
2.2.1. The actual scale interval (d) and the verification scale interval (e) shall be in the form:
1 × 10k, 2 × 10k or 5 × 10k mass units,
k being any integer or zero.
2.2.2. For all instruments other than those with auxiliary indicating devices:
d = e
2.2.3. For instruments with auxiliary indicating devices the following conditions apply:
e = 1 × 10k g
d < e ≤ 10 d
except for instruments of class I with d <10−4 g, for which e = 10−3 g.
3.1. Instruments with one weighing range
Instruments equipped with an auxiliary indicating device shall belong to class I or class II. For these instruments the minimum capacity lower limits for these two classes are obtained from Table 1 by replacement in column 3 of the verification scale interval (e) by the actual interval (d).
If d < 10−4 g, the maximum capacity of class I may be less than 50 000 e.
3.2. Multiple weighing ranges are permitted, provided they are clearly indicated on the instument. Each individual weighing range is classified according to 3.1. If the weighing ranges fall into different accuracy classes the instrument shall comply with the severest of the requirements that apply for the accuracy classes in which the weighing ranges fall.
3.3.1. Instruments with one weighing range may have several partial weighing ranges (multi-interval instruments).
Multi-interval instruments shall not be equipped with an auxiliary indicating device.
3.3.2. Each partial weighing range i of multi-interval instruments is defined by:
its verification scale interval ei with e(i + 1) > ei
its maximum capacity Maxi with Maxr = Max
its minimum capacity Mini with Mini = Max(i − 1) and Min1 = Min
where:
i = 1, 2, … r,
i = partial weighing range number,
r = the total number of partial weighing ranges.
All capacities are capacities of net load, irrespective of the value of any tare used.
3.3.3. The partial ranges are classified according to Table 2. All partial weighing ranges shall fall into the same accuracy class, this class being the instrument’s accuracy class.
i = 1, 2, … ri = partial weighing range numberr = total number of partial weighing ranges
Class | Verification scale interval (e) | Minimum capacity (Min) | Number of verification scale intervals | |
---|---|---|---|---|
Minimum Value | Minimum value(1) | Maximum value | ||
(1) For i=r the corresponding column of Table 1 applies, with e replaced by er. | ||||
I | 0,001 g ≤ ei | 100 e1 | 50 000 | — |
II | 0,001 g ≤ ei ≤ 0,05 g | 20 e1 | 5 000 | 100 000 |
0,1 g ≤ ei | 50 e1 ; | 5 000 | 100 000 | |
III | 0,1 g ≤ ei | 20 e1 | 500 | 10 000 |
IIII | 5 g ≤ ei | 10 e1 | 50 | 1 000 |
4.1. On implementation of the procedures laid down in Article 8, the error of indication shall not exceed the maximum permissible error of indication as shown in Table 3. In case of digital indication the error of indication shall be corrected for the rounding error.
The maximum permissible errors apply to the net and tare value for all possible loads, excluded preset tare values.
Load | Maximum permissible error | |||
---|---|---|---|---|
Class I | Class II | Class III | Class IIII | |
0 ≤m ≤ 50 000 e | 0 ≤m ≤ 5 000 e | 0 ≤m ≤ 500 e | 0 ≤m ≤ 50 e | ± 0,5 e |
50 000 e < m ≤ 200 000 e | 5 000 e < m ≤ 20 000 e | 500 e < m ≤ 2 000 e | 50 e < m ≤ 200 e | ± 1,0 e |
200 000 e < m | 20 000 e < m ≤ 100 000 e | 2 000 e < m ≤ 10 000 e | 200 e < m ≤ 1 000 e | ± 1,5 e |
4.2. The maximum permissible errors in service are twice the maximum permissible errors fixed in section 4.1.
5. Weighing results of an instrument shall be repeatable, and shall be reproducible by the other indicating devices used and with other methods of balancing used.
The weighing results shall be sufficiently insensitive to changes in the position of the load on the load receptor.
6. The instrument shall react to small variations in the load.
7.1. Instruments of classes II, III and IIII, liable to be used in a tilted position, shall be sufficiently insensitive to the degree of tilting that can exist in a normal installed condition.
7.2. The instruments shall meet the metrological requirements within the temperature range specified by the manufacturer. The value of this range shall be at least equal to:
5°C for an instrument in class I,
15°C for an instrument in class II,
30°C for an instrument in class III or IIII.
In the absence of a manufacturer’s specification, the temperature range of −10°C to +40°C applies.
7.3. Instruments operated from a mains power supply shall meet the metrological requirements under conditions of power supply within the limits of normal fluctuation.
Instruments operated from battery power shall indicate whenever the voltage drops below the minimum required value and shall under those circumstances either continue to function correctly or be automatically put out of service.
7.4. Electronic instruments, except those in class I and in class II if e is less than 1 g, shall meet the metrological requirements under conditions of high relative humidity at the upper limit of their temperature range.
7.5. Loading an instrument in class II, III or IIII for a prolonged period of time shall have a negligible influence on the indication at load or on the zero indication immediately after removal of the load.
7.6. Under other conditions the instruments shall either continue to function correctly or be automatically put out of service.
OJ No. L39, 15.2.1980, p. 40.
OJ No. L34, 9.2.2000, p. 17.
Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.
Original (As Enacted or Made): The original version of the legislation as it stood when it was enacted or made. No changes have been applied to the text.
Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:
Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:
Click 'View More' or select 'More Resources' tab for additional information including: