4.—(1) The Department shall only use sampling bottles which—
(a)have been—
(i)sterilised in an autoclave for at least 15 minutes at 121 degrees Celsius;
(ii)dry sterilised at no lower than 160 degrees Celsius and no higher than 170 degrees Celsius for at least one hour; or
(iii)irradiated by their manufacturer and not used previously;
(b)are of a size which allows sufficient water to be taken and analysed for the presence of Intestinal enterococci and Escherichia coli; and
(c)are made of transparent and colourless material.
(2) The Department shall—
(a)use aseptic techniques to maintain the sterility of the sample bottles; and
(b)clearly identify every sample taken by marking in indelible ink the sample bottle and associated paperwork.