Directive 2004/10/EC of the European Parliament and of the CouncilShow full title

Directive 2004/10/EC of the European Parliament and of the Council of 11 February 2004 on the harmonisation of laws, regulations and administrative provisions relating to the application of the principles of good laboratory practice and the verification of their applications for tests on chemical substances (codified version) (Text with EEA relevance)

4.Apparatus, material, and reagentsU.K.

1.Apparatus, including validated computerised systems, used for the generation, storage and retrieval of data, and for controlling environmental factors relevant to the study should be suitably located and of appropriate design and adequate capacity.U.K.
2.Apparatus used in a study should be periodically inspected, cleaned, maintained, and calibrated according to standard operating procedures. Records of these activities should be maintained. Calibration should, where appropriate, be traceable to national or international standards of measurement.U.K.
3.Apparatus and materials used in a study should not interfere adversely with the test systems.U.K.
4.Chemicals, reagents, and solutions should be labelled to indicate identity (with concentration if appropriate), expiry date and specific storage instructions. Information concerning source, preparation date and stability should be available. The expiry date may be extended on the basis of documented evaluation or analysis.U.K.