All reagents must be of analytical quality.
Into a tared 25 ml volumetric flask, introduce about 650 mg of chloroform (5.1.1). Accurately reweigh the flask and contents. Make up to 25 ml with 95 % ethanol (5.1.3). Weigh and calculate the percentage by mass of chloroform in this solution.
Make up in a similar manner to the chloroform reference solution but weigh 270 mg of 2,4-dimethylheptane (5.1.2) into the 25 ml volumetric flask.
Into a 100 ml tared transfer bottle, purged or evacuated according to the procedure described in 5.4 of Chapter II of the abovementioned Directive, introduce about 5 ml of either of the internal standard solutions (5.1.5 or 5.1.6). Use a 10 or 20 ml glass syringe, without needle, adapted to the transfer piece following the technique described in paragraph 5 of Chapter II of the above-mentioned Commission Directive. Reweigh to determine the quantity introduced. Using the same technique, transfer into this bottle about 50 g of the contents of the aerosol dispenser sample. Again reweigh to determine the quantity of sample transferred. Mix well.
Inject about 10 μl using the specified microsyringe (5.2.2). Make five injections.
Into a 50 ml volumetric flask, accurately weigh about 500 mg of nitromethane (5.1.4) and either 500 mg of chloroform (5.1.1) or 210 mg of 2,4-dimethylheptane (5.1.2). Make up to volume with 95 % ethanol (5.1.3). Mix well. Place 5 ml of this solution into a 20 mg volumetric flask. Make up to volume with 95 % ethanol (5.1.3).
Inject about 10 μl using the specified microsyringe (5.2.2). Make five injections.
This is in two parts, the first containing didecyl phthalate on Gas Chrom Q as packing, the second having Ucon 50 HB 280X on Gas Chrom Q as packing. The prepared combined column must yield a resolution ‘R’ equal to, or better than, 1,5, where:
let:
=
retention times (in minutes),
=
peak widths at half height (in millimetres),
=
the chart speed (in millimetres per minute).
As examples the following two parts yield the required resolution:
Column A:
Material: stainless steel.
Length: 1,5 m.
Diameter: 3 mm.
Packing: 20 % didecyl phthalate on Gas Chrom Q (100 to 120 mesh).
Column B:
Materia: stainless steel.
Length: 1,5 m.
Diameter: 3 mm.
Packing: 20 % Ucon 50 HB 280X on Gas Chrom Q (100 to 120 mesh).
A suitable sensitivity setting for the electrometer of the flame ionization detector is 8 × l0-10A.
The following have been found suitable:
Injection port: 150 oC,
Detector: 150 oC,
Column: between 50 and 80 oC depending upon individual columns and apparatus.
Carrier gas: nitrogen.
Pressure: 2,1 bar.
Flow: 40 ml/min
Detector supplies: as specified by the makers of the detector.