Commission Decisionof 12 August 2002implementing Council Directive 96/23/EC concerning the performance of analytical methods and the interpretation of results(notified under document number C(2002) 3044)(Text with EEA relevance)(2002/657/EC)

 Help about what version

What Version

 Help about advanced features

Advanced Features

 Help about UK-EU Regulation

Legislation originating from the EU

When the UK left the EU, legislation.gov.uk published EU legislation that had been published by the EU up to IP completion day (31 December 2020 11.00 p.m.). On legislation.gov.uk, these items of legislation are kept up-to-date with any amendments made by the UK since then.

Close

This item of legislation originated from the EU

Legislation.gov.uk publishes the UK version. EUR-Lex publishes the EU version. The EU Exit Web Archive holds a snapshot of EUR-Lex’s version from IP completion day (31 December 2020 11.00 p.m.).

Changes to legislation:

There are currently no known outstanding effects by UK legislation for Commission Decisionof 12 August 2002implementing Council Directive 96/23/EC concerning the performance of analytical methods and the interpretation of results(notified under document number C(2002) 3044)(Text with EEA relevance)(2002/657/EC), 3.. Help about Changes to Legislation

Close

Changes to Legislation

Revised legislation carried on this site may not be fully up to date. At the current time any known changes or effects made by subsequent legislation have been applied to the text of the legislation you are viewing by the editorial team. Please see ‘Frequently Asked Questions’ for details regarding the timescales for which new effects are identified and recorded on this site.

3.U.K.VALIDATION

Validation shall demonstrate that the analytical method complies with the criteria applicable for the relevant performance characteristics.

Different control purposes require different categories of methods. The following table determines which performance characteristic shall be verified for which type of method.

Table 9

Classification of analytical methods by the performance characteristics that have to be determined

Detection limit CCßDecision limit CCαTrueness/recoveryPrecisionSelectivity/specificityApplicability/ruggedness/stability
S = screening methods; C = confirmatory methods; + = determination is mandatory.
Qualitative methodsS+++
C++++
Quantitative methodsS++++
C++++++

3.1.U.K.VALIDATION PROCEDURES

This chapter provides examples and/or references for validation procedures of analytical methods. Other approaches to demonstrate that the analytical method complies with performance criteria for the performance characteristics may be used, provided that they achieve the same level and quality of information.

Validation can also be performed by conducting an interlaboratory study such as established by Codex Alimentarius, ISO or the IUPAC (12), or according to alternative methods such as single laboratory studies or in-house validation (13)(14). This part focuses on single laboratory studies (on in-house validation) using a modular approach. This approach consists of:

1.

a set of common performance characteristics independent of the validation model used and

2.

more specific model-dependent procedures as described in Table 10.

Table 10

Model-independent and model-dependent performance parameters

Validation
Model-independent performance parametersModel-dependent performance parameters
Common performance characteristics (3.1.1)Conventional validation approach (3.1.2)In-house validation approach (3.1.3)
SpecificityRecoveryRecovery
TruenessRepeatabilityRepeatability
Ruggedness: minor changesWithin-laboratory reproducibilityWithin-laboratory reproducibility
StabilityReproducibilityReproducibility
Decision limit (CC α)Decision limit (CC α)
Detection capability (CCβ)Detection capability (CCβ)
Calibration curvesCalibration curve
Ruggedness: major changesRuggedness

3.1.1Model-independent performance characteristicsU.K.

Irrespective of the validation approach chosen, the following performance characteristics have to be determined. To minimise the workload, a carefully designed and statistically sound approach can be used to combine experiments performed to determine different parameters.

3.1.1.1SpecificityU.K.

For analytical methods, the power of discrimination between the analyte and closely related substances (isomers, metabolites, degradation products, endogenous substances, matrix constituents, etc) is important. Two approaches are necessary to check for interferences.

Therefore, potentially interfering substances shall be chosen and relevant blank samples shall be analysed to detect the presence of possible interferences and to estimate the effect of the interferences:

  • select a range of chemically related compounds (metabolites, derivatives, etc.) or other substances likely to be encountered with the compound of interest that may be present in the samples;

  • analyse an appropriate number of representative blank samples (n ≥ 20) and check for any interferences (signals, peaks, ion traces) in the region of interest where the target analyte is expected to elute;

  • additionally, representative blank samples shall be fortified at a relevant concentration with substances that are likely to interfere with the identification and/or quantification of the analyte;

  • after analysis, investigate whether:

    • the presence may lead to a false identification,

    • the identification of the target analyte is hindered by the presence of one or more of the interferences, or

    • the quantification is influenced notably.

3.1.1.2TruenessU.K.

In this paragraph, the determination of trueness (one component of accuracy) is described. Trueness can only be established by means of certified reference material (CRM). A CRM be used whenever available. The procedure is described in detail in ISO 5725-4 (5). An example is given below:

  • analyse six replicates of the CRM in accordance with the test instructions for the method,

  • determine the concentration of the analyte present in each sample of the replicates,

  • calculate the mean, the standard deviation and the coefficient of variation (%) for these concentrations,

  • calculate the trueness by dividing the detected mean concentration by the certified value (measured as concentration) and multiply by 100, to express the result as a percentage.

Trueness ( %) = mean recovery-corrected concentration detected × 100/certified value.

If no CRM is available, instead of trueness, the recovery can be determined as described under 4.1.2.1 below.

3.1.1.3Applicability/ruggedness (minor changes)U.K.

Such studies use the deliberate introduction of minor reasonable variations by the laboratory and the observation of their consequences.

The pre-investigative studies have to be carried out by selecting factors of the sample pre-treatment, clean up and analysis, which may influence the measurement results. Such factors may include the analyst, the source and the age of reagents, solvents, standards and sample extracts, the rate of heating, the temperature, the pH-value as well as many other factors that may occur in the laboratory. These factors should be modified in an order of magnitude that matches the deviations usually encountered among laboratories.

  • Identify possible factors that could influence the results.

  • Vary each factor slightly.

  • Conduct a ruggedness test using the approach of Youden (15)(16). (Other approved methods may be used at this point. The Youden approach, however, keeps the required time and effort to a minimum). The Youden approach is a fractional factorial design. Interactions between the different factors cannot be detected.

  • Where a factor is found to influence the measurement results significantly, conduct further experiments to decide on the acceptability limits of this factor.

  • Factors that significantly influence the results should be identified clearly in the method protocol.

The basic idea is not to study one alteration at a time but to introduce several variations at once. As an example, let A, B, C, D, E, F, G denote the nominal values for seven different factors that could influence the results, if their nominal values are changed slightly. Let their alternative values be denoted by the corresponding lower case letters a, b, c, d, e, f and g. This results in 27 or 128 different possible combinations.

It is possible to choose a subset of eight of these combinations that have a balance between capital and small letters (Table 11). Eight determinations have to be made, which will use a combination of the chosen factors (A-G). The results of the determinations are shown in Table 11 below as S-Z.

Table 11
Experiment design for ruggedness studies (minor changes)
Factor value FCombination of determinations number
12345678
For calculations see examples for ruggedness testing in 3.3.
A/aAAAAaaaa
B/bBBbbBBbb
C/cCcCcCcCc
D/dDDddddDD
E/eEeEeeEeE
F/fFffFFffF
G/gGggGgGGg
Observed result RSTUVWXYZ

3.1.1.4StabilityU.K.

It has been observed that insufficient stability of the analyte or matrix constituents in the sample during storage or analysis may give rise to significant deviations in the outcome of the result of analysis. Furthermore, the stability of the calibration standard in solution should be checked. Usually analyte stability is well characterised under various storage conditions. Monitoring of the storage condition will form part of the normal laboratory accreditation system. When this is not known, examples are given below on how the stability can be determined.

Stability of the analyte in solution:

  • Prepare fresh stock solutions of the analyte(s) and dilute as specified in the test instructions to yield sufficient aliquots (e.g. 40) of each selected concentration (around the minimum required performance limit for substances for which no permitted limit has been established or around the permitted limit for other substances. Prepare both solutions of the analyte used for fortification and used in the final analysis solution, and any other solution that is of interest (e.g. derivatised standards).

  • Measure the analyte content in the freshly prepared solution according to the test instructions.

  • Dispense appropriate volumes into suitable containers, label and store according to the scheme:

    Table 12
    Scheme for the determination of analyte stability in solution
    – 20 °C+ 4 °C+ 20 °C
    Dark10 aliquots10 aliquots10 aliquots
    Light10 aliquots
  • The storing time could be selected as one, two, three and four weeks or longer if necessary, e.g. until the first degradation phenomena are observable during identification and/or quantification. The maximum storing time and the optimum storing conditions have to be recorded.

  • The calculation of the concentration of the analyte(s) in each aliquot should be carried out by using the solution of the analyte freshly prepared at the time of analysis as 100 %.

    • Analyte Remaining (%) = Ci × 100/Cfresh

    • Ci= concentration at time point

    • Cfresh= concentration of fresh solution

Stability of analyte(s) in matrix

  • Whenever possible, incurred samples should be used. When no incurred material is available, matrix fortified with the analyte should be used.

  • When incurred material is available, the concentration in the material should be determined while the material is still fresh. Further aliquots of material could be taken after one, two, four and 20 weeks and the concentrations should be determined. The tissue should be stored at least minus 20 °C or lower if required.

  • If no incurred material is available, take some blank material and homogenise it. Divide the material into five aliquots. Fortify each aliquot with the analyte, which should preferably be prepared in a small quantity of aqueous solution. Analyse one aliquot immediately. Store the remaining aliquots at least minus 20 °C or lower if required and analyse them after one, two, four and 20 weeks.

3.1.1.5Calibration curvesU.K.

When calibration curves are used for quantification:

  • at least five levels (including zero) should be used in the construction of the curve,

  • the working range of the curve should be described,

  • the mathematical formula of the curve and the goodness-of-fit of the data to the curve should be described,

  • acceptability ranges for the parameters of the curve should be described.

When serial calibration based on a standard solution is necessary, acceptable ranges shall be indicated for the parameters of the calibration curve, which may vary from series to series.

3.1.2Conventional validation proceduresU.K.

The calculation of the parameters in accordance with conventional methods requires the performance of several individual experiments. Each performance characteristic has to be determined for each major change (see under applicability/ruggedness above). For multi-analyte methods, several analytes can be analysed simultaneously, as long as possibly relevant interferences are ruled out previously. Several performance characteristics can be determined in a similar way. So, to minimise workload, it is advised to combine experiments as much as possible (e.g., repeatability and within-laboratory reproducibility with specificity, analysis of blank samples to determine the decision limit and testing for specificity).

3.1.2.1RecoveryU.K.

If there is no CRM available, the recovery has to be determined by experiments using fortified blank matrix using, for example, the following scheme:

  • select 18 aliquots of a blank material and fortify six aliquots at each of 1, 1,5 and 2 times the minimum required performance limit or 0,5, 1 and 1,5 times the permitted limit,

  • analyse the samples and calculate the concentration present in each sample,

  • using the equation below, calculate the recovery for each sample,

  • calculate the mean recovery and CV from the six results at each level,

  • % Recovery = 100 × measured content/fortification level.

This conventional method for the determination of recovery is a variant of the standard addition method described in 3.5, when:

  • the sample is considered as a blank sample instead of a sample to be analysed,

  • it is considered that yield (1) and recovery (2) are similar for the two test portions,

  • the test samples have the same masses and the test portion extracts the same volumes,

  • the amount of the calibration standard that is added to the second (spiked) test portion is noted xADD. (xADD = ρA.VA),

  • x1 is the measured value for the blank and x2 the measured value for the second (spiked) test portion,

  • then, % Recovery = 100 (x2 - x1)/xADD.

When any of the above conditions is (or is supposed) not to be achieved, then the complete procedure for determination of the recovery by mean of the standard addition method as described in 3.5 has to be performed.

3.1.2.2RepeatabilityU.K.

  • Prepare a set of samples of identical matrices, fortified with the analyte to yield concentrations equivalent to 1, 1,5 and 2 times the minimum required performance limit or 0,5, 1 and 1,5 times the permitted limit.

  • At each level the analysis should be performed with at least six replicates.

  • Analyse the samples.

  • Calculate the concentration detected in each sample.

  • Find the mean concentration, standard deviation and the coefficient of variation (%) of the fortified samples.

  • Repeat these steps on at least two other occasions.

  • Calculate the overall mean concentrations and CVs for the fortified samples.

3.1.2.3Within-laboratory reproducibilityU.K.

  • Prepare a set of samples of specified test material (identical or different matrices), fortified with the analyte(s) to yield concentrations equivalent to 1, 1,5 and 2 times the minimum required performance limit or 0,5, 1 and 1,5 times the permitted limit.

  • At each level the analysis should be performed with at least six replicates.

  • Repeat these steps on at least two other occasions with different operators and different environmental conditions, e.g. different batches of reagents, solvents etc., different room temperatures, different instruments, etc. if possible.

  • Analyse the samples.

  • Calculate the concentration detected in each sample.

  • Find the mean concentration, standard deviation and the coefficient of variation ( %) of the fortified samples.

3.1.2.4ReproducibilityU.K.

When reproducibility has to be verified, laboratories should participate in collaborative studies according to ISO 5725-2 (5).

3.1.2.5Decision Limit (CCɑ)U.K.

The decision limit has to be established according to the requirements for identification or identification plus quantification as defined under ‘Performance criteria and other requirements for analytical methods’ (part 2).

In the case of substances for which no permitted limit has been established, CCɑ can be established:

  • either by the calibration curve procedure according to ISO 11843 (17) (here referred to as critical value of the net state variable). In this case blank material shall be used, which is fortified at and above the minimum required performance level in equidistant steps. Analyse the samples. After identification, plot the signal against the added concentration. The corresponding concentration at the y-intercept plus 2,33 times the standard deviation of the within-laboratory reproducibility of the intercept equals the decision limit. This is applicable to quantitative assays only (ɑ = 1 %),

  • or by analysing at least 20 blank materials per matrix to be able to calculate the signal to noise ratio at the time window in which the analyte is expected. Three times the signal to noise ratio can be used as decision limit. This is applicable to quantitative and qualitative assays.

In the case of substances an with established permitted limit, CCɑ can be established:

  • either by the calibration curve procedure according to ISO 11843 (17) (here referred to as critical value of the net state variable). In this case blank material shall be used, which is fortified around the permitted limit in equidistant steps. Analyse the samples. After identification, plot the signal against the added concentration. The corresponding concentration at the permitted limit plus 1,64 times the standard deviation of the within-laboratory reproducibility equals the decision limit (ɑ = 5 %),

  • or by analysing at least 20 blank materials per matrix fortified with the analyte(s) at the permitted limit. The concentration at the permitted limit plus 1,64 times the corresponding standard deviation equal the decision limit (ɑ = 5 %).

See also Article 5 and point 3.2.

3.1.2.6Detection capability (CCβ)U.K.

The detection capability should be determined according to the requirements for screening, identification or identification plus quantification as defined (see part 2).

In the case of substances for which no permitted limit has been established, CCβ can be established by:

  • the calibration curve procedure according to ISO 11843 (17) (here referred to as minimum detectable value of the net state variable). In this case representative blank material shall be used, which is fortified at and below the minimum required performance level in equidistant steps. Analyse the samples. After identification, plot the signal against the added concentration. The corresponding concentration at the decision limit plus 1,64 times the standard deviation of the within-laboratory reproducibility of the mean measured content at the decision limit equals the detection capability (β = 5 %),

  • analysing at least 20 blank materials per matrix fortified with the analyte(s) at the decision limit. Analyse the samples and identify the analytes. The value of the decision limit plus 1,64 times the standard deviation of the within-laboratory reproducibility of the measured content equals the detection capability (β = 5 %),

  • where no quantitative results are available, the detection capability can be determined by the investigation of fortified blank material at and above the decision limit. In this case the concentration level, where only ≤ 5 % false compliant results remain, equals the detection capability of the method. Therefore, at least 20 investigations for at least one concentration level have to be carried out in order to ensure a reliable basis for this determination.

In the case of substances for which a permitted limit has been established, CCβ can be established:

  • either by the calibration curve procedure according to ISO 11843 (17) (here referred to as minimum detectable value of the net state variable). In this case representative blank material shall be used, which is fortified around the permitted limit in equidistant steps. Analyse the samples and identify the analyte(s). Calculate the standard deviation of the mean measured content at the decision limit. The corresponding concentration at the value of the decision limit plus 1,64 times the standard deviation of the within-laboratory reproducibility equals the detection capability (β = 5 %),

  • or by analysing at least 20 blank materials per matrix fortified with the analyte(s) at the decision limit. The value of the decision limit plus 1,64 times the corresponding standard deviation equals the detection capability (β = 5 %).

See also section 3.2.

3.1.2.7Ruggedness (major changes)U.K.

The analytical method should be tested under different experimental conditions, which include for example, different species, different matrices or different sampling conditions. The changes introduced should be major. The importance of these changes can be evaluated, for instance, using the Youden approach (15)(16). Each performance characteristic should be determined for all major changes that have been shown to have a significant effect on the performance of the assay.

3.1.3Validation according to alternative modelsU.K.

When alternative validation procedures are applied, the underlying model and strategy with the respective prerequisites, assumptions and formulae shall be laid down in the validation protocol or at least references shall be given to their availability. In the following an example for an alternative approach is given. When applying e.g. the in-house validation model, the performance characteristics are determined in a manner that permits validation for major changes within the same validation procedure. This requires design of an experimental plan for validation.

3.1.3.1Experimental planU.K.

An experimental plan has to be designed depending on the number of different species and different factors under investigation. Hence, the first step of the entire validation procedure shall consider the sample populations that will be analysed in the laboratory in the future in order to select the most important species and those factors which may influence the measurement results. Subsequently, the concentration range shall be chosen in a purpose-adapted way according to the level of interest.

Example:

  • several analytes can be investigated simultaneously with the analytical method being validated,

  • two variations of the leading factor have been identified (A and B). Leading factors form the basis on which the factor levels are combined. These leading factors may include factors such as species or matrix. In this example the leading factor was varied on two levels, i.e. two different species (species A and B) were considered. In general, it is possible to vary the leading factors on more than two levels, which only increase the number of analyses to be performed,

  • the selected factors are to be varied on two levels (indicated as either + or -).

Table 13
Examples for factors considered important for a validation procedure
Gender of the animal(factor 1)
Breed(factor 2)
Transport conditions(factor 3)
Storing conditions(factor 4)
Freshness of the sample(factor 5)
Fattening conditions(factor 6)
Different operators with different experience(factor 7).
Table 14
Possible experimental plan for the above example
Species Factor 1Factor 2Factor 3Factor 4Factor 5Factor 6Factor 7Sample No
A+++++1
A+++2
A+++3
A+++++4
A+++++5
A+++6
A+++7
A+8
B++++++9
B++++10
B++++11
B++12
B++13
B++++14
B++++15
B++16

As each sample (each factor level combination) has to be spiked with four different concentrations around the level of interest, and one blank sample shall be analysed for each level, 5 x 16 = 80 analyses have to be performed for the entire validation experiment.

From these 80 measurement results it is possible to compute (13)(14).

Recovery

  • repeatability per concentration level (sir),

  • within-laboratory reproducibility per concentration level (sir),

  • decision limit (CCɑ),

  • detection capability (CCβ),

  • power curve (β-error rate versus concentration (see 3.1.3.2),

  • ruggedness of major changes; ruggedness to minor changes can be determined according to paragraph 3.1.1.3,

  • 16 sample-related calibration curves,

  • one overall calibration curve,

  • prediction interval of the overall calibration curve,

  • matrix-induced deviations (smat),

  • run-induced deviations (srun),

  • effect of the individual factors on the measurement results.

These performance characteristics allow the comprehensive evaluation of the performance level of the method, since not only the influence of the individual factors is investigated, but also the relevant combinations of these factors. With the help of this experiment design it is possible to decide if one or the other of the selected factors shall be excluded from the overall calibration curve, because it significantly deviates from the standard deviations of the other factors.

3.1.3.2Power curveU.K.

The power curve provides information about the detection capability of the method within the chosen concentration range. It refers to the β-error risk when applying the investigated method. The power curve allows to calculate the detection capabilities for the respective categories (screening, confirmation) or types (qualitative or quantitative) of methods for a certain β-error (e.g. 5 %).

Figure 1

Power curve

Figure 1 shows an example of the graphical establishment of detection capability (CCβ) of an analytical method. This particular method has a remaining risk of taking a false decision of 5 % at a concentration of 0,50 μg/kg. At a concentration of 0,55 μg/kg the risk of taking a false compliant decision decreases to 1 %.

3.1.3.3ReproducibilityU.K.

The determination of a method's reproducibility by the single laboratory studies (in-house validation) concept requires repeated participation in proficiency studies in accordance with ISO guide 43-1 (3) and 43-2 (4). The laboratories are allowed to choose their own methods, provided these methods are used under routine conditions. The standard deviation of the laboratory can be used to assess the reproducibility of the method.

3.2.U.K.GRAPHICAL REPRESENTATION OF THE DIFFERENT ANALYTICAL LIMITS

Figure 2

Substances for which no permitted limit has been established

Figure 3

Substances with an established permitted limit

3.3.U.K.CALCULATION EXAMPLE FOR RUGGEDNESS TESTING OF MINOR CHANGES ACCORDING TO THE YOUDEN APPROACH (16)

Comparison of averages (A)

AA = Σ(Ai)/4

Compare the averages of the capital letters (AA to AG) with the averages of their corresponding small letters (Aa to Ag). If a factor have an effect, the difference will be significant larger than the differences of the other factors.

A robust method should not be affected by changes encountered almost certainly between laboratories.

If there is no outstanding difference, the most realistic measure of the random error is given by the seven differences.

AB = Σ(Bi)/4
AC = Σ(Ci)/4
AD = Σ(Di)/4
AE = Σ(Ei)/4
AF = Σ(Ei)/4
AG = Σ(Gi)/4
Aa = Σ(ai)/4
Ab = Σ(bi)/4
Ac = Σ(ci)/4
Ad = Σ(di)/4
Ae = Σ(ei)/4
Af = Σ(fi)/4
Ag = Σ(gi)/4
Differences (Di)Squares of differences (Di2)
Da = A – a = Σ(Ai) – Σ(ai)Da2 = value a
Db = B – b = Σ(Bi) – Σ(bi)Db2 = value b
Dc = C – c = Σ(Ci) – Σ(ci)Dc2 = value c
Dd = D – d = Σ(Di) – Σ(di)Dd2 = value d
De = E – e = Σ(Ei) – Σ(ei)De2 = value e
Df = F – f = Σ(Fi) – Σ(fi)Df2 = value f
Dg = G – g = Σ(Gi) – Σ(gi)Dg2 = value g

Standard deviation of the differences Di (SDi):U.K.

When SDi is significantly larger than the standard deviation of the method carried out under within-laboratory reproducibility conditions according (see above) it is a foregone conclusion that all factors together have an effect on the result even if every single factor does not show a significant influence and that the method is not sufficiently robust against the chosen modifications.

3.4.U.K.CALCULATION EXAMPLES FOR THE IN-HOUSE VALIDATION PROCEDURE

Examples and calculations for the in-house validation protocol as described under validation according to alternative models (3.1.3) (13) (14).

3.5.U.K.EXAMPLES FOR THE STANDARD ADDITION METHOD

A test sample with a content T of the analyte is divided in two test portions 1 and 2 of respective masses m1 and m2. The test portion 2 is spiked with a volume VA of a solution of concentration ρA of the analyte. Two extracts of the test portions of respective volumes V1 and V2 are obtained after extraction and purification steps of the method. The recovery of the analyte is supposed to be rc. Both extracts are assayed with a measurement method of sensitivity b and give an analytical response of x1 and x2 respectively.

If assumed that rc and b are the same for the analyte in the native sample and in the spiked sample, then the content T can be calculated as:

The method will allow the determination of the recovery rc. Then, in addition with the assay described above, part of the extract of the test portion 1 (volume V3) is spiked with a known amount ρB.VB of the analyte and assayed. The analytical response is x3 and the recovery is:

Moreover, it is possible to calculate the sensitivity b, as:

All conditions of application and details have been described (18).

(1)

Yield: that fraction of mass of the analyte contained in the sample, which is present in the final extract.

(2)

Recovery (here): that fraction of mass of the analyte added to the sample, which is present in the final extract. Throughout the rest of the document it is assumed that yield and recovery are equal and therefore only the term ‘recovery’ is used.

Back to top

Options/Help

Print Options

Close

Legislation is available in different versions:

Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.

Original (As adopted by EU): The original version of the legislation as it stood when it was first adopted in the EU. No changes have been applied to the text.

Close

See additional information alongside the content

Geographical Extent: Indicates the geographical area that this provision applies to. For further information see ‘Frequently Asked Questions’.

Show Timeline of Changes: See how this legislation has or could change over time. Turning this feature on will show extra navigation options to go to these specific points in time. Return to the latest available version by using the controls above in the What Version box.

Close

Opening Options

Different options to open legislation in order to view more content on screen at once

Close

More Resources

Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:

  • the original print PDF of the as adopted version that was used for the EU Official Journal
  • lists of changes made by and/or affecting this legislation item
  • all formats of all associated documents
  • correction slips
  • links to related legislation and further information resources
Close

Timeline of Changes

This timeline shows the different versions taken from EUR-Lex before exit day and during the implementation period as well as any subsequent versions created after the implementation period as a result of changes made by UK legislation.

The dates for the EU versions are taken from the document dates on EUR-Lex and may not always coincide with when the changes came into force for the document.

For any versions created after the implementation period as a result of changes made by UK legislation the date will coincide with the earliest date on which the change (e.g an insertion, a repeal or a substitution) that was applied came into force. For further information see our guide to revised legislation on Understanding Legislation.

Close

More Resources

Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:

  • the original print PDF of the as adopted version that was used for the print copy
  • correction slips

Click 'View More' or select 'More Resources' tab for additional information including:

  • lists of changes made by and/or affecting this legislation item
  • confers power and blanket amendment details
  • all formats of all associated documents
  • links to related legislation and further information resources