9.—(1) An automatic discontinuous totaliser shall have no feature likely to facilitate fraudulent use, whereas possibilities for unintentional misuse shall be minimal.
(2) An automatic discontinuous totaliser shall be suitable for its intended use taking account of the practical working conditions and shall not require unreasonable demands of the user in order to obtain a correct measurement result.
(3) Where an automatic discontinuous totaliser is designed for the measurement of values of the measurand that are constant over time, the instrument shall be insensitive to small fluctuations of the value of the measurand, or shall take appropriate action.
(4) An automatic discontinuous totaliser shall be robust and its materials of construction shall be suitable for the conditions in which it is intended to be used.
(5) An automatic discontinuous totaliser shall be designed so as to allow the control of the measuring tasks after the instrument has been placed on the market and put into use. If necessary, special equipment or software for this control shall be part of the instrument. The test procedure shall be described in the operation manual.
(6) When an instrument has associated software which provides other functions besides the measuring function, the software that is critical for the metrological characteristics shall be identifiable and shall not be inadmissibly influenced by the associated software.
(7) Means shall be provided to limit the effects of tilt, loading and rate of operation such that MPEs are not exceeded in normal operation.
(8) Adequate material handling facilities shall be provided to enable the instrument to respect the MPE during normal operation.
(9) Any operator control interface shall be clear and effective.
(10) The integrity of the display (where present) shall be verifiable by the operator.
(11) Adequate zero setting capability shall be provided to enable the instrument to respect the MPE during normal operation.
(12) Any result outside the measurement range shall be identified as such, where a printout is possible.
Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.
Original (As Enacted or Made):The original version of the legislation as it stood when it was enacted or made. No changes have been applied to the text.
Explanatory Memorandum sets out a brief statement of the purpose of a Statutory Instrument and provides information about its policy objective and policy implications. They aim to make the Statutory Instrument accessible to readers who are not legally qualified accompany any Statutory Instrument or Draft Statutory Instrument laid before Parliament from June 2004 onwards.
Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:
Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:
Click 'View More' or select 'More Resources' tab for additional information including: