xmlns:atom="http://www.w3.org/2005/Atom"

SCHEDULE 2ANNEX I TO COUNCIL DIRECTIVE 90/384/EEC of 20TH JUNE 1990 ON THE HARMONISATION OF THE LAWS OF MEMBER STATES RELATING TO NON-AUTOMATIC WEIGHING INSTRUMENTS

METROLOGICAL REQUIREMENTS

Units of mass

1.  The units of mass used shall be the legal units within the meaning of Directive 80/181/EEC(1) as last amended by Directive 1999/103/EC(2).

Accuracy classes

The following accuracy classes have been defined:

Scale intervals

2.2.1.  The actual scale interval (d) and the verification scale interval (e) shall be in the form:

2.2.2.  For all instruments other than those with auxiliary indicating devices:

d = e

2.2.3.  For instruments with auxiliary indicating devices the following conditions apply:

Classification

3.1.  Instruments with one weighing range

Instruments with multiple weighing ranges

3.2.  Multiple weighing ranges are permitted, provided they are clearly indicated on the instument. Each individual weighing range is classified according to 3.1. If the weighing ranges fall into different accuracy classes the instrument shall comply with the severest of the requirements that apply for the accuracy classes in which the weighing ranges fall.

Multi-interval instruments

3.3.1.  Instruments with one weighing range may have several partial weighing ranges (multi-interval instruments).

Multi-interval instruments shall not be equipped with an auxiliary indicating device.

3.3.2.  Each partial weighing range i of multi-interval instruments is defined by:

3.3.3.  The partial ranges are classified according to Table 2. All partial weighing ranges shall fall into the same accuracy class, this class being the instrument’s accuracy class.

TABLE 2
Multi-interval instruments

i = 1, 2, … ri = partial weighing range numberr = total number of partial weighing ranges

ClassVerification scale interval (e)Minimum capacity (Min)Number of verification scale intervals
Minimum Value

Minimum value(1)

Maximum value

(1)

For i=r the corresponding column of Table 1 applies, with e replaced by er.

I0,001 g ≤ ei100 e150 000
II0,001 g ≤ ei ≤ 0,05 g20 e15 000100 000
0,1 g ≤ ei50 e1 ;5 000100 000
III0,1 g ≤ ei20 e150010 000
IIII5 g ≤ ei10 e1501 000

Accuracy

4.1.  On implementation of the procedures laid down in Article 8, the error of indication shall not exceed the maximum permissible error of indication as shown in Table 3. In case of digital indication the error of indication shall be corrected for the rounding error.

The maximum permissible errors apply to the net and tare value for all possible loads, excluded preset tare values.

TABLE 3
Maximum permissible errors
LoadMaximum permissible error
Class IClass IIClass IIIClass IIII
0 ≤m ≤ 50 000 e0 ≤m ≤ 5 000 e0 ≤m ≤ 500 e0 ≤m ≤ 50 e± 0,5 e
50 000 e < m ≤ 200 000 e5 000 e < m ≤ 20 000 e500 e < m ≤ 2 000 e50 e < m ≤ 200 e± 1,0 e
200 000 e < m20 000 e < m ≤ 100 000 e2 000 e < m ≤ 10 000 e200 e < m ≤ 1 000 e± 1,5 e

4.2.  The maximum permissible errors in service are twice the maximum permissible errors fixed in section 4.1.

5.  Weighing results of an instrument shall be repeatable, and shall be reproducible by the other indicating devices used and with other methods of balancing used.

The weighing results shall be sufficiently insensitive to changes in the position of the load on the load receptor.

6.  The instrument shall react to small variations in the load.

Influence quantities and time

7.1.  Instruments of classes II, III and IIII, liable to be used in a tilted position, shall be sufficiently insensitive to the degree of tilting that can exist in a normal installed condition.

7.2.  The instruments shall meet the metrological requirements within the temperature range specified by the manufacturer. The value of this range shall be at least equal to:

In the absence of a manufacturer’s specification, the temperature range of −10°C to +40°C applies.

7.3.  Instruments operated from a mains power supply shall meet the metrological requirements under conditions of power supply within the limits of normal fluctuation.

Instruments operated from battery power shall indicate whenever the voltage drops below the minimum required value and shall under those circumstances either continue to function correctly or be automatically put out of service.

7.4.  Electronic instruments, except those in class I and in class II if e is less than 1 g, shall meet the metrological requirements under conditions of high relative humidity at the upper limit of their temperature range.

7.5.  Loading an instrument in class II, III or IIII for a prolonged period of time shall have a negligible influence on the indication at load or on the zero indication immediately after removal of the load.

7.6.  Under other conditions the instruments shall either continue to function correctly or be automatically put out of service.

(1)

OJ No. L39, 15.2.1980, p. 40.

(2)

OJ No. L34, 9.2.2000, p. 17.