xmlns:atom="http://www.w3.org/2005/Atom"

Regulations 2(1) and (2),18(7)(b) and 23(1)(a)

SCHEDULE 2ANNEX I TO COUNCIL DIRECTIVE 90/384/EEC of 20TH JUNE 1990 ON THE HARMONISATION OF THE LAWS OF MEMBER STATES RELATING TO NON-AUTOMATIC WEIGHING INSTRUMENTS

The essential requirements that must be met by the instruments referred to in Article 1(2)(a) are set out below. The terminology used is that of the Organisation Internationale de Métrologie Légale(1).

Preliminary observation

Where an instrument includes or is connected to more than one indicating or printing device used for the applications listed in Article 1(2)(a), those devices which repeat the results of the weighing operation and which cannot influence the correct functioning of the instrument shall not be subject to the essential requirements if the weighing results are printed or recorded correctly and indelibly by a part of the instrument which meets the essential requirements and the results are accessible to both parties concerned by the measurement. However, in the case of instruments used for direct sales to the public, display and printing devices for the vendor and the customer must fulfil the essential requirements.

METROLOGICAL REQUIREMENTS

Units of mass

1.  The units of mass used shall be the legal units within the meaning of Directive 80/181/EEC(2) as last amended by Directive 1999/103/EC(3).

Accuracy classes

The following accuracy classes have been defined:

Scale intervals

2.2.1.  The actual scale interval (d) and the verification scale interval (e) shall be in the form:

2.2.2.  For all instruments other than those with auxiliary indicating devices:

d = e

2.2.3.  For instruments with auxiliary indicating devices the following conditions apply:

Classification

3.1.  Instruments with one weighing range

Instruments with multiple weighing ranges

3.2.  Multiple weighing ranges are permitted, provided they are clearly indicated on the instument. Each individual weighing range is classified according to 3.1. If the weighing ranges fall into different accuracy classes the instrument shall comply with the severest of the requirements that apply for the accuracy classes in which the weighing ranges fall.

Multi-interval instruments

3.3.1.  Instruments with one weighing range may have several partial weighing ranges (multi-interval instruments).

Multi-interval instruments shall not be equipped with an auxiliary indicating device.

3.3.2.  Each partial weighing range i of multi-interval instruments is defined by:

3.3.3.  The partial ranges are classified according to Table 2. All partial weighing ranges shall fall into the same accuracy class, this class being the instrument’s accuracy class.

TABLE 2
Multi-interval instruments

i = 1, 2, … ri = partial weighing range numberr = total number of partial weighing ranges

ClassVerification scale interval (e)Minimum capacity (Min)Number of verification scale intervals
Minimum Value

Minimum value(1)

Maximum value

(1)

For i=r the corresponding column of Table 1 applies, with e replaced by er.

I0,001 g ≤ ei100 e150 000
II0,001 g ≤ ei ≤ 0,05 g20 e15 000100 000
0,1 g ≤ ei50 e1 ;5 000100 000
III0,1 g ≤ ei20 e150010 000
IIII5 g ≤ ei10 e1501 000

Accuracy

4.1.  On implementation of the procedures laid down in Article 8, the error of indication shall not exceed the maximum permissible error of indication as shown in Table 3. In case of digital indication the error of indication shall be corrected for the rounding error.

The maximum permissible errors apply to the net and tare value for all possible loads, excluded preset tare values.

TABLE 3
Maximum permissible errors
LoadMaximum permissible error
Class IClass IIClass IIIClass IIII
0 ≤m ≤ 50 000 e0 ≤m ≤ 5 000 e0 ≤m ≤ 500 e0 ≤m ≤ 50 e± 0,5 e
50 000 e < m ≤ 200 000 e5 000 e < m ≤ 20 000 e500 e < m ≤ 2 000 e50 e < m ≤ 200 e± 1,0 e
200 000 e < m20 000 e < m ≤ 100 000 e2 000 e < m ≤ 10 000 e200 e < m ≤ 1 000 e± 1,5 e

4.2.  The maximum permissible errors in service are twice the maximum permissible errors fixed in section 4.1.

5.  Weighing results of an instrument shall be repeatable, and shall be reproducible by the other indicating devices used and with other methods of balancing used.

The weighing results shall be sufficiently insensitive to changes in the position of the load on the load receptor.

6.  The instrument shall react to small variations in the load.

Influence quantities and time

7.1.  Instruments of classes II, III and IIII, liable to be used in a tilted position, shall be sufficiently insensitive to the degree of tilting that can exist in a normal installed condition.

7.2.  The instruments shall meet the metrological requirements within the temperature range specified by the manufacturer. The value of this range shall be at least equal to:

In the absence of a manufacturer’s specification, the temperature range of −10°C to +40°C applies.

7.3.  Instruments operated from a mains power supply shall meet the metrological requirements under conditions of power supply within the limits of normal fluctuation.

Instruments operated from battery power shall indicate whenever the voltage drops below the minimum required value and shall under those circumstances either continue to function correctly or be automatically put out of service.

7.4.  Electronic instruments, except those in class I and in class II if e is less than 1 g, shall meet the metrological requirements under conditions of high relative humidity at the upper limit of their temperature range.

7.5.  Loading an instrument in class II, III or IIII for a prolonged period of time shall have a negligible influence on the indication at load or on the zero indication immediately after removal of the load.

7.6.  Under other conditions the instruments shall either continue to function correctly or be automatically put out of service.

Design and Construction

General requirements

8.1.  Design and construction of the instruments shall be such that the instruments will preserve their metrological qualities when properly used and installed, and when used in an environment for which they are intended. The value of the mass must be indicated.

8.2.  When exposed to disturbances, electronic instruments shall not display the effects of significant faults, or shall automatically detect and indicate them.

Upon automatic detection of a significant fault, electronic instruments shall provide a visual or audible alarm that shall continue until the user takes corrective action or the fault disappears.

8.3.  The requirements of 8.1 and 8.2 shall be met on a lasting basis during a period of time that is normal in view of the intended use of such instruments.

8.4.  When external equipment is connected to an electronic instrument through an appropriate interface the metrological qualities of the instrument shall not be adversely influenced.

8.5.  The instruments shall have no characteristics likely to facilitate fraudulent use, whereas possibilities for unintentional misuse shall be minimal. Components that may not be dismantled or adjusted by the user shall be secured against such actions.

8.6.  Instruments shall be designed to permit ready execution of the statutory controls laid down by this Directive.

Indication of weighing results and other weight values

9.  The indication of the weighing results and other weight values shall be accurate, unambiguous and non-misleading and the indicating device shall permit easy reading of the indication under normal conditions of use.

Printing of weighing results and other weight values

10.  Printed results shall be correct, suitably identified and unambiguous. The printing shall be clear, legible, non-erasable and durable.

Levelling

11.  When appropriate, instruments shall be fitted with a levelling device and a level indicator, sufficiently sensitive to allow proper installation.

Zeroing

12.  Instruments may be equipped with zeroing devices. The operation of these devices shall result in accurate zeroing and shall not cause incorrect measuring results.

Tare devices and preset tare devices

13.  The instruments may have one or more tare devices and a preset tare device. The operation of the tare devices shall result in accurate zeroing and shall ensure correct net weighing. The operation of the preset tare device shall ensure correct determination of the calculated net value.

Instruments for direct sales to the public with a maximum capacity not greater than 100 kg: additional requirements

14.  Instruments for direct sale to the public shall show all essential information about the weighing operation and, in the case of price-indicating instruments, shall clearly show the customer the price calculation of the product to be purchased.

Price labelling instruments

15.  Price labelling instruments shall meet the requirements of price indicating instruments for direct sale to the public, as far as applicable to the instrument in question. The printing of a price label shall be impossible below a minimum capacity.

(1)

See “Non-automatic weighing instruments Part I: Metrological and Technical requirements—Tests” OIML R 76-1 Edition 1992 (E) published by the International Organisation for Legal Metrology.

(2)

OJ No. L39, 15.2.1980, p. 40.

(3)

OJ No. L34, 9.2.2000, p. 17.

(4)

OJ No. L39, 15.2.1980, p. 40.