xmlns:atom="http://www.w3.org/2005/Atom" xmlns:atom="http://www.w3.org/2005/Atom"

[F1ANNEX IV U.K. COMMUNITY ANALYSIS METHOD FOR NEUTRAL ALCOHOL

I. GENERAL U.K.

For the purposes of this Annex:

(a)

the repeatability limit is the value below which the absolute difference between two single test results obtained under the same conditions (same operator, same apparatus, same laboratory and a short interval of time) may be expected to lie with a specified probability;

(b)

the reproducibility limit is the value below which the absolute difference between two single test results obtained under different conditions (different operators, different apparatus and/or different laboratories and/or different time) may be expected to lie with a specified probability.

The term single test result means the value obtained when the standardised test method is applied fully once to a single sample. Unless otherwise stated, the probability shall be 95 %.

II. METHODS U.K.

Introduction U.K.

1. PREPARATION OF THE SAMPLE FOR ANALYSIS U.K.
1.1. General U.K.

The volume of laboratory sample intended for analysis must normally be 1,5 l unless a larger quantity is required for a specific determination.

1.2. Sample preparation U.K.

The sample shall be made homogeneous before analysis.

1.3. Preservation U.K.

The prepared sample shall always be kept in an air-tight and moisture-tight container and stored so that deterioration is prevented; in particular seals of cork, rubber and plastic should not come into direct contact with the alcohol and the use of sealing wax is expressly prohibited.

2. REAGENTS U.K.
2.1. Water U.K.
2.1.1. Wherever mention is made of water for solution, dilution or washing purposes, dilution or washing purposes, distilled water, or demineralised water of a least equivalent purity shall be used. U.K.
2.1.2. Wherever reference is made to solution or dilution , without further indication of a reagent, an aqueous solution is intended. U.K.
2.2. Chemicals U.K.

All chemicals shall be of analytical reagent quality except where otherwise specified.

3. EQUIPMENT U.K.
3.1. List of equipment U.K.

The list of equipment contains only those items with a specialised use and items with a particular specification.

3.2. Analytical balance U.K.

Analytical balance means a balance with a sensitivity of 0,1 mg or better.

4. EXPRESSION OF RESULTS U.K.
4.1. Results U.K.

The result stated in the analytical report is the mean value obtained from at least two determinations, the repeatability (r) of which is satisfactory.

4.2. Calculation of results U.K.

Except where otherwise specified, the results shall be calculated as g per hl of ethanol at 100 % vol.

4.3. Number of significant figures U.K.

The result shall not contain more significant figures than are justified by the precision of the method of analysis used.

Method 1: Determination of alcohol content U.K.

The alcoholic strength by volume of the alcohol shall be determined in accordance with existing national provisions or, in the event of dispute, by means of alcoholometers or hydrometers as defined in Council Directive 76/765/EEC of 27 July 1976 on the approximation of the laws of the Member States relating to alcoholometers and alcohol hydrometers (1) .

It shall be expressed as percentage by volume, as prescribed by Council Directive 76/766/EEC of 27 July 1976 on the approximation of the laws of the Member States relating to alcohol tables (2) .

Method 2: Evaluation of colour and/or clarity U.K.

1. SCOPE AND FIELD OF APPLICATION U.K.

The method enables the colour and/or clarity of neutral alcohol to be assessed.

2. DEFINITION U.K.

The colour and/or clarity: the colour and/or clarity as assessed by the procedure specified.

3. PRINCIPLE U.K.

The colour and clarity are assessed visually by comparison with water against a white background and a black background respectively.

4. APPARATUS U.K.

Glass cylinders, colourless, at least 40 cm in height.

5. PROCEDURE U.K.

Place two glass cylinders (4) on the white background or black background and fill one cylinder with sample to a depth of approximately 40 cm and the other with water to the same depth.

Observe the sample from above i.e. through the length of the cylinder, and compare it with the comparison cylinder.

6. INTERPRETATION U.K.

Assess the colour and/or clarity of the sample when observed as set out in 5.

Method 3: Determination of permanganate clearing time U.K.

1. SCOPE AND FIELD OF APPLICATION U.K.

The method determines the permanganate clearing time of neutral alcohol.

2. DEFINITION U.K.

The permanganate clearing time, as determined by the method specified, is the number of minutes required for the colour of the sample to match that of the colour standard after adding 1 ml of a 1 mmol/l potassium permanganate solution to 10 ml of the sample.

3. PRINCIPLE U.K.

The time for the colour of the sample, after addition of potassium permanganate, to match that of a colour standard is determined and defined as the permanganate clearing time.

4. REAGENTS U.K.
4.1. Potassium permanganate solution, 1 mmol/l. Prepare immediately before use. U.K.
4.2. Colour solution A (red) U.K.
4.3. Colour solution B (yellow) U.K.
4.4. Colour standard solution U.K.

Pipette 13 ml of colour solution A and 5,5 ml colour solution B into a 100 ml volumetric flask and make up to the mark with water at 20  o C.

Note: U.K.

Colour solutions A and B may be stored in the ark at 4  o C for several months; the colour standard should be freshly prepared from time to time. U.K.

5. APPARATUS U.K.
5.1. 100 ml Nessler tubles of colourless transparent glass, graduated to 50 ml, with ground-glass stopper, or test tubes, colourless, circa 20 mm diameter. U.K.
5.2. Pipettes, 1, 2, 5, 10 and 50 ml. U.K.
5.3. Thermometer, range up to 50  o C in 0,1 or 0,2  o C. U.K.
5.4. Analytical balance. U.K.
5.5. Water bath, thermostatically controlled at 20 ± 0,5  o C. U.K.
5.6. Volumetric flasks, 100 and 1 000  ml with ground glass stoppers. U.K.
6. PROCEDURE U.K.
6.1.
Note: U.K.

Take care not to expose the sample solution to direct sunlight during the test. U.K.

7. EXPRESSION OF RESULTS U.K.
7.1. Interpretation of the time of clearing shall be the time required for the colour of the sample tube to match that of the standard tube. U.K.

For a neutral alcohol that time must be at least 18 minutes at a temperature of 20  o C.

7.2. Repeatability U.K.

The difference in the times of two tests, carried out simultaneously or in rapid succession, by the same analyst, on the same sample, under the same conditions, shall not exceed two minutes.

8. NOTES U.K.
8.1. Traces of manganese dioxide have a catalysing effect on the reaction, ensure that pipettes and test tubes are used which have been scrupulously cleaned and reserved exclusively for the purpose. Clean them with hydrochloric acid and rinse thoroughly with water, no brown coloration shall be visible on the glass apparatus. U.K.
8.2. The quality of the water to prepare the dilute permanganate solution (4.1) shall be carefully monitored; it must not consume any permanganate. If the required quality cannot be obtained, distilled water should be brought to the boil and a small quantity of permanganate added in order to obtain a very slightly pink coloration. This should then be cooled and used for the dilution. U.K.
8.3. For some samples the decoloration may occur without passing through the exact shade of the reference solution. U.K.
8.4. The permanganate test may be distorted when the alcohol sample for analysis has not been stored in a perfectly clean glass flask, sealed with either a ground glass stopper which has been rinsed with alcohol or another stopper encased in tin or aluminium. U.K.

Method 4: Determination of aldehydes U.K.

1. SCOPE AND FIELD OF APPLICATION U.K.

The method determines aldehydes, expressed as acetaldehyde, in neutral alcohol.

2. DEFINITION U.K.

The aldehyde content: the content of aldehydes, expressed as acetaldehyde, as determined by the method specified.

3. PRINCIPLE U.K.

The colour obtained after the reaction of the sample with Schiff's reagent is compared with standard solutions having a known acetaldehyde content.

4. REAGENTS U.K.

p-rosaniline hydrochloride (basic fuchsin)

sodium sulphite or anhydrous sodium metabisulphite

hydrochloric acid, density p 20  = 1.19 g/ml

powdered active carbon

starch solution, prepared from 1 g soluble starch and 5 mg Hgl 2 (preservative) which are suspended in a little cold water, mixed with 500 ml boiling water, boiled for 5 minutes and filtered when cold.

iodine solution, 0,05 mol/l

1-amino-ethanol CH 3 .CH(NH 2 )OH (MW 61.08)

Schiff's reagent

Notes: U.K.
(1)

The Schiff's reagent should be prepared at least 14 days before it is used.

(2)

The free SO 2 content in the reagent should be between 2,8 and 6,0 mmol/100 ml, the pH must be 1.

Determination of free SO 2

  • Pipette 10 ml Schiff's reagent into a 250 ml Erlenmeyer flask,

  • add 200 ml water,

  • add 5 ml starch solution,

  • titrate with 0,05 mol/l iodine solution to starch end-point,

if the free SO 2 content is outside the indicated range it should either be:

  • raised with a calculated quantity of sodium metabisulphite (0,126 g Na 2 SO 3 /100 ml reagent per mmol SO 2 lacking), or

  • lowered by bubbling air through the reagent.

Calculation of free SO 2 in the reagent.

mmol free SO 2 /100 ml reagent:

Important: U.K.

It other methods are used to prepare the Schiff's reagent, the sensitivity of the reagent should be checked so that during the test: U.K.

  • there is no colouration with the aldehyde-free reference alcohol,

  • the pink colouration should be visible from 0,1 g acetaldehyde per hl alcohol at 100 % vol.

(3)

Purification of commercial 1-amino-ethanol

  • Dissolve 5 g 1-amino-ethanol completely in circa 15 ml absolute ethanol,

  • add circa 50 ml dry diethyl ether (1-amino-ethanol precipitates),

  • leave for several hours in a refrigerator,

  • filter off the crystals and wash with dry diethyl ether,

  • dry for three to four hours in a desiccator over sulphuric acid in a partial vacuum.

Note: U.K.

The cleaned 1-araino-ethanol must be white; if not, repeat the recrystallisation process. U.K.

5. APPARATUS U.K.
5.1. Colorimetric tubes, each provided with a ground-glass stopper, capacity 20 ml. U.K.
5.2. 1 ml, 2 ml, 3 ml, 4 ml, 5 ml and 10 ml pipettes. U.K.
5.3. Water bath, thermostatically controlled at 20 ± 0,5  o C. U.K.
5.4. Spectrophotometer with cuvettes of 50 mm path length. U.K.
6. PROCEDURE U.K.
6.1. Preliminary remark U.K.

When using this method to determine aldehyde content ensure that the sample's alcohol content is at least 90,0 % vol. If not it must be raised by adding corresponding amounts of aldehyde-free ethanol.

6.2. Calibration curve U.K.
6.3. Determining the aldehyde content U.K.
Notes: U.K.
(1)

To determine the aldehyde figures it is necessary to check the validity of the calibration curves by comparison with test solution; if not, the calibration curve must be prepared again.

(2)

Ensure that the blank is always colourless.

7. EXPRESSION OF RESULTS U.K.
7.1. Formula and method of calculation U.K.

Construct a graph of optical density against concentration of acetaldehyde and determine the concentration in the sample by reference to this plot.

The content of aldehydes, expressed as acetaldehyde, in g/hl ethanol at 100 % vol is given by

where:

A

=

the content, in g per hl of acetaldehyde in the sample solution as determined by reference to the standard curve,

T

=

the alcoholic strength by volume of the sample as determined by method 1.

7.2. Repeatability U.K.

The difference between the results of two determinations, carried out simultaneously or in rapid concession, by the same analyst, on the same sample, under the same conditions, shall not exceed 0,1 g aldehyde per hl Ethanol at 100 % vol.

Method 5: Determination of higher alcohols U.K.

1. SCOPE AND FIELD OF APPLICATION U.K.

The method determines budget alcohols, expressed as 2 methylpropan-1-ol, in neutral alcohol.

2. DEFINITION U.K.

The higher alcohol content: the content of higher alcohols, expressed as 2-methylpropan-l-ol as determined by the method specified.

3. PRINCIPLE U.K.

The absorbance of the coloured products resulting from the reaction of higher alcohols and an aromatic aldehyde in hot dilute sulphuric acid (Komarowsky reaction) are determined at 560 nm, corrected for the presence of any aldehyde in the sample and then compared with that produced by 2-methylpropan-1-ol reacting under the same conditions.

4. REAGENTS U.K.
4.1. Salicyl aldehyde solution, 1 % mas. Prepare by adding 1 g of salicyl aldehyde to 99 g of ethanol at 96 % vol (which is free of fusel oil). U.K.
4.2. Sulphuric acid, concentrated, density 1,84 g/m 2 . U.K.
4.3. 2-methylpropan-1-ol. U.K.
4.4. Standard 2-methylpropan-1-ol solutions U.K.

Dilute 2-methylpropan-1-ol (4.3) with an aqueous solution of 96 % vol ethanol to give a series of standards containing 0,1, 0,2, 0,4, 0,6 and 1,0 g of 2-methylpropan-1-ol per hl of solution.

4.5. Standard acetaldehyde solutions. U.K.

Prepare the standard acetaldehyde solutions as described in section 6.2 of method 4.

4.6. Ethanol 96 % vol free from higher alcohols and aldehydes. U.K.
5. APPARATUS U.K.
5.1. UV-VIS spectrophotometer, capable of determining absorbance of solutions at 560 nm. U.K.
5.2. Spectrophotometer cuvettes, 10, 20 and 50 mm pathlengths. U.K.
5.3. Water bath, thermostatically regulated at 20 ± 0,5  o C. U.K.
5.4. Colorimetric tubes made of thick glass Pyrex or similar with ground-glass stoppers, content circa 50 ml. U.K.
6. PROCEDURE U.K.
6.1. Aldehyde content U.K.

Determine the aldehyde content, expressed as acetaldehyde, in the sample using methanol 4.

6.2. Calibration curve: 2-methylpropan-1-ol U.K.

Pipette 10 ml of each of the 2-methylpropan-1-ol standards (4.4) into 50 ml glass cylinders each fitted with ground glass stoppers. Pipette 1 ml of the salicyl aldehyde solution (4.1) into the cylinders and then 20 ml of sulphuric acid (4.2). Mix the contents thoroughly by carefully tilting the cylinders backwards and forwards several times (care being taken to lift the stopper in occasionally). Leave for 10 minutes at room temperature and then put in the water bath (5.3) at 20 ± 0,5  o C. After 20 minutes pour the contents into a series of spectrophotometer cuvettes.

Exactly 30 minutes after adding the sulphuric acid determine the absorbance of the solutions at 560 nm using water in the reference cuvette of the spectrophotometer.

Construct a calibration curve of absorbance against 2-methylpropan-1-ol concentration.

6.3. Calibration curve — aldehydes U.K.

Repeat 6.2 but replacing the 10 ml of each of the 2-methylpropan-1-ol standards by 10 ml of each of the acetaldehyde standards.

Construct a calibration curve of absorbance at 560 nm against acetaldehyde concentration.

6.4. Sample determination U.K.

Repeat 6.2 but replacing the 10 ml of the 2-methylpropan-1-ol standards by 10 ml of the sample.

Determine the absorbance of the sample.

7. EXPRESSION OF RESULTS U.K.
7.1. Formula and method of calculation U.K.
7.1.1. Correct the absorbance of the sample by subtracting the value of the absorbance corresponding to the aldehyde concentration in the sample (obtained from the calibration curve constructed under 6.3). U.K.
7.1.2. Determine the concentration of higher alcohols, expressed as 2-methalpropan-1-ol, in the sample from the calibration curve constructed under 6.2 but using the correct absorbance (7.1.1). U.K.
7.1.3. The concentration of higher alcohols, expressed as 2-methylpropan-1-ol in g per hl ethanol at 100 % vol, is given by: U.K.

where:

A

=

the concentration of higher alcohols in the sample as calculated in 7.1.2.

T

=

the alcoholic strength by volume of the sample as determined by method 1.

7.2. Repeatability U.K.

The difference between the results of two determinations, when carried out simultaneously or in rapid succession, by the same analyst, on the same sample, under the same conditions, shall not exceed 0,2 g per hl ethanol at 100 % vol.

Method 6: Determination of total acidity U.K.

1. SCOPE AND FIELD OF APPLICATION U.K.

The method determines the total acidity, expressed as acetic acid, of neutral alcohol.

2. DEFINITION U.K.

The total acidity content, expressed as acetic acid: the content of total acidity, expressed as acetic acid, as determined by the method specified.

3. PRINCIPLE U.K.

The sample, after de-gassing, is titrated against standard sodium hydroxide solution and the acidity calculated as acetic acid.

4. REAGENTS U.K.
4.1. Sodium hydroxide solutions, 0,01 mol/l and 0,1 mol/l, stored so that contact with carbon dioxide is minimised. U.K.
4.2. Indigo carmine solution (A) U.K.

Phenol red solution (B)

5. APPARATUS U.K.
5.1. Burette or automatic titrater. U.K.
5.2. Pipette, 100 ml. U.K.
5.3. Round-bottom flask with ground-glass stopper, 250 ml. U.K.
5.4. Reflux condenser with ground-glass stopper. U.K.
6. PROCEDURE U.K.
7. EXPRESSION OF RESULTS U.K.
7.1. Formula and method of calculation U.K.

The content of total acidity, expressed as acetic acid, in g per hl ethanol at 100 % vol is given by:

where:

V

=

the number of ml of 0,01 mol/l sodium hydroxide required for neutralisation.

T

=

the alcoholic strength by volume of the sample as determined by method 1.

7.2. Repeatability U.K.

The difference between the results of two determinations, carried out simultaneously or in rapid succession, by the same analyst, on the same sample, under the same conditions, shall not exceed 0,1 g per hl ethanol at 100 % vol.

Method 7: Determination of esters U.K.

1. SCOPE AND FIELD OF APPLICATION U.K.

The method determines esters, expressed as ethyl acetate, in neutral alcohol.

2. DEFINITION U.K.

The ester content: the content of esters, expressed as ethyl acetate, as determined by the method specified.

3. PRINCIPLE U.K.

Esters react quantitatively with hydroxylamine hydrochloride in alkaline solution to form hydroxylamic acids. These then form coloured complexes with ferric ions in acid solution. The optical densities of these complexes are measures at 525 nm.

4. REAGENTS U.K.
4.1. Hydrochloric acid 4 mol/l. U.K.
4.2. Ferric chloride solution, 0,37 mol/l in 1 mol/l hydrochloric acid. U.K.
4.3. Hydroxylamine Hydrochloride, 2 mol/l. Store in a refrigerator. U.K.
4.4. Sodium hydroxide solution, 3,5 mol/l. U.K.
4.5. Ethyl acetate standard solutions containing 0,0, 0,2, 0,4, 0,6, 0,8 and 1,0 g ethyl acetate per hl ester-free ethanol at 96 % vol. U.K.
5. APPARATUS U.K.
5.1. Spectrophotometer with 50 mm pathlength cuvettes. U.K.
6. PROCEDURE U.K.
6.1. Calibration curve U.K.
6.2. Determination of ester content U.K.
7. EXPRESSION OF RESULTS U.K.
7.1. Formula and method of calculation U.K.

Plot the optical densities of the standards against their concentrations.

The ester content (expressed as ethyl acetate = A) corresponding to the absorbence value is read off on the graph and calculated according to the formula:

and given in 9 per hl ethanol at 100 % vol,

where T

=

alcohol content of the sample in % vol determination as described in method 1.

7.2. Repeatability U.K.

The difference between the results of two determinations, carried out simultaneously or in rapid succession, by the same analyst, on the same sample, under the same conditions, should not exceed 0,1 g esters, as ethyl acetate, per hl ethanol at 100 % vol.

Method 8: Determination of volatile nitrogen bases U.K.

1. SCOPE AND FIELD OF APPLICATION U.K.

The method determines volatile nitrogen bases, expressed as nitrogen, in neutral alcohol.

2. DEFINITION U.K.

The volatile nitrogen bases content: the content of volatile nitrogen bases, expressed as nitrogen, as determined by the method specified.

3. PRINCIPLE U.K.

The sample is evaporated to a small volume in the presence of sulphuric acid and the ammonia content then determined using the conway micro-diffusion technique.

4. REAGENTS U.K.
4.1. Suphuric acid, 1 mol/l. U.K.
4.2. Boric acid indicator solution. Dissolve 10 g of boric acid, 8 mg of bromocresol green and 4 mg of methyl red in 30 % vol propan-2-ol and make up to 1 000  ml with 30 % vol propan-2-ol. U.K.
4.3. Potasium hydroxide solution, 500 g/l; carbon dioxide free. U.K.
4.4. Hydrochloric acid, 0,02 mol/l. U.K.
5. APPARATUS U.K.
5.1. Evaporating dish, of sufficient capacity to accept 50 ml of sample. U.K.
5.2. Water bath. U.K.
5.3. Conway flask with tightly fitting lid; see figure 1 for description and suggested dimensions. U.K.
5.4. Micro-burette, capacity 2 to 5 ml, graduated in 0,01 ml. U.K.
6. PROCEDURE U.K.
6.1. Pipette 50 mls of the sample (with an anticipated nitrogen content of less than 0,2 g per hl sample take 200 ml of sample) into a glass dish, add 1 ml of 1 mol/l sulphuric acid (4.1), place the dish (5.1) on a water bath (5.2) and evaporate until there is about 1 ml remaining. U.K.
6.2. Pipette 1 ml of the boric acid indicator solution (4.2) into the inner chamber of the Conway flask (5.3) and wash the residue liquid from the evaporation process (6.1) into the outer chamber. Slightly tilt the Conway flask and add about 1 ml of the potassium hydroxide solution (4.3) to the outer chamber as quickly as possible but as far from the majority of the liquid in the outer chamber as possible. Immediately seal the Conway flask by covering with a tightly fitting lid smeared with grease. U.K.
6.3. Mix the two solutions in the outer chamber taking care that there is no liquid spillage from either chamber to the other. Allow to stand for two hours. U.K.
6.4. Titrate the ammonia in the inner chamber against 0,02 mol/l hydrochloric acid (4.4) using a micro-burette (5.4) to neutralisation. The volume of acid used should be between 0,2 and 0,9 ml; let the volume of acid used by V 1  ml. U.K.
6.5. Carry out a blank titration by repeating sections 6.1 to 6.4 but replacing the 50 ml of sample in section 6.1 by the same volume of water. Let the volume of hydrochloric acid used be V 2  ml. U.K.
7. EXPRESSION OF RESULTS U.K.
7.1. Formula and method of calculation. U.K.

The content of volable nitrogen bases, in g per hl ethanol at 100 % vol, calculated and expressed as nitrogen, is given by:

Where

V 1

=

the volume, to ml, at the hydrochloric acid used to neutralize the sample.

V 2

=

the volume, in ml, of hydrochloric acid used in the blank test.

T

=

the alcoholic strength by volume of the sample as determined by method 1.

E

=

the quantity of sample used in ml.

7.2. Repeatability U.K.

The difference between the results of two determinations, carried out simultaneously or in rapid succession, by the same analyst, on the same sample, under the same conditions, shall not exceed 0,05 g per hl ethanol at 100 % vol.

Method 9: Determination of methanol U.K.

1. SCOPE AND FIELD OF APPLICATION U.K.

The method determines the content of methanol in neutral alcohol.

2. DEFINITION U.K.

The methanol content: the content of methanol as determined by the method specified.

3. PRINCIPLE U.K.

The methanol concentration is determined by direct injection of sample into a gas liquid chromatography apparatus.

4. PROCEDURE U.K.

Any GLC method is suitable provided that the gas chromatographic column and conditions employed are capable of achieving a clear separation between methanol, acetaldehyde, ethanol and ethyl acetate. The limit of detection of methanol in ethanol shall be less than 2 g/hl.

5. REPEATABILITY U.K.

The difference between the results of two determinations, carried out simultaneously or in rapid succession, by the same anylst, on the same sample, under the same conditions, shall not exceed 2 g methanol per hl ethanol at 100 % vol.

Method 10: Determination of dry residue U.K.

1. SCOPE AND FIELD OF APPLICATION U.K.

The method determines the dry residue content of neutral alcohol.

2. DEFINITION U.K.

The dry residue content: the dry matter content as determined by the method specified.

3. PRINCIPLE U.K.

An aliquot of the sample is dried at 103  o C and the residue determined gravimetrically.

4. APPARATUS U.K.
4.1. Water bath, boiling. U.K.
4.2. Evaporating dish of suitable capacity. U.K.
4.3. Desiccator, containing freshly activated silica gel (or an equivalent desiccant) with a moisture content indicator. U.K.
4.4. Analytical balance. U.K.
4.5. Oven, thermostatically controlled at 103 ± 2  o C. U.K.
5. PROCEDURE U.K.

Accurately weigh, to the nearest 0,1 mg, a clean dry evaporating dish (4.2) (m o ). Pipette in several operations if necessary a suitable volume of sample into the dish (100 — 250 ml) (V o ml). Place the dish with sample on the boiling water bath (4.1) and allow to. dry. Place in the oven (4.5) at 103 ± 2  o C for 30 minutes and then transfer dish with residue into a desiccator (4.3). Allow the dish to cool for 30 minutes and then weigh, to the nearest 0,1 mg, the dish with residue (M 1 ).

6. EXPRESSION OF RESULTS U.K.
6.1. Formula and method of calculation U.K.

The content of dry residue, to g per hl of ethanol at 100 % vol is given by:

where:

M o

=

the mass, in g, of the clean dry dish,

M 1

=

the mass, in g, of the dish and residue after drying,

V o

=

the volume of sample taken for drying, and

T

=

the alcohols strength by volume of the sample as determined by method 1.

6.2. Repeatability U.K.

The difference between the results of two determinations, carried out simultaneously or in rapid succession, by the same analyst, on the same sample, under the same conditions, shall not exceed 0,5 g per hl of ethanol at 100 % vol.

Method 11: Limit test for the absence of furfural U.K.

1. SCOPE AND FIELD OF APPLICATION U.K.

The method detects furfural in neutral alcohol.

2. DEFINITION U.K.

The detection of the limit tests concentration of furfural: the limit test result as determined by the method specified.

3. PRINCIPLE U.K.

The alcohol sample is mixed with aniline and glacial acetic acid. The presence of furfural is indicated by a salmon pink colour appearing in the solution within 20 minutes of mixing.

4. REAGENTS U.K.
4.1. Aniline, freshly distilled. U.K.
4.2. Acetic acid, glacial. U.K.
5. APPARATUS U.K.

Tubes, fitted with ground glass stoppers.

6. PROCEDURE U.K.

Pipette 10 ml of the sample into a tube (5); add 0,5 ml of aniline and 2 ml of glacial acetic acid. Shake the tube and contents to mix.

7. EXPRESSION OF RESULTS U.K.
7.1. Limit test interpretation U.K.

If the time of development of any salmon pink colouration in the tube is less than 20 minutes the test is positive and the sample contains furfural.

7.2. Observations U.K.

The results of two limit tests, carried out simultaneously or in rapid succession, by the same analyst, on the same sample, under the same conditions, shall be identical.

Method 12: UV test U.K.

1. SCOPE U.K.

This method determines the optical transparence of neutral alcohol.

2. PRINCIPLE U.K.

The optical transparence of the sample in the wavelength range 220 to 270 nm is measured against a defined reference substance of high optical transparence.

3. APPARATUS U.K.
3.1. UV-VIS spectrophotometer U.K.
3.2. Quartz cuvettes, 10 mn path length, same spectrum transmission. U.K.
4. REAGENTS U.K.

n-Hexane for spectroscopy.

5. PROCEDURE U.K.
6. EXPRESSION OF RESULTS U.K.

The absorbence values found at 270, 240, 230 and 220 nm may not exceed the following figures: 0,02, 0,08, 0,18 and 0,3.

The absorbence curve must be smooth and regular.

Method 13: Determination of 14 C content in ethanol U.K.

1. METHOD TO DETERMINE TYPE OF ALCOHOL U.K.

Determination of the 14 C content in ethanol permits a distinction to be made between alcohol from fossil fuels (synthesis alcohol) and alcohol from recent raw materials (fermentation alcohol).

2. DEFINITION U.K.

The 14 C content of ethanol is understood to be the 14 C content determined using the method described here.

The natural 14 C content in the atmosphere (the reference value), which is absorbed by living vegetation by assimilation, is not a constant value. The reference value is therefore determined on ethanol from raw materials of the most recent vegetation period. This annual reference value is determined each year by collaborative analyses organised by the Community Bureau of References and the Joint Research Centre, Ispra

3. PRINCIPLE U.K.

The 14 C content of samples containing alcohol with at least 85 % mass ethanol is determined directly by liquid scintillation count.

4. REAGENTS U.K.
4.1. Toluene scintillator U.K.

5,0 g 2,5-diphenyloxazole (PPO)

0,5 g p-bis-[4-methyl-5-phenyloxazolyl(2)]-benzene (dimethyl-POPOP) in 1 litre analytical grade toluene.

Commercial, ready-to-use toluene scintillators of this composition can also be used.

4.2. 14 C standard U.K.

n-Hexadecane 14 C with an activity of about 1 × 10 6 dpm/g (approximately 1,67 × 10 6 cBq/g) and a guaranteed accuracy of determined activity of ± 2 % rel.

4.3. 14 C-free ethanol U.K.

Synthesis alcohol from raw materials of fossil origin with at least 85 % mass ethanol, to determine the background.

4.4. Alcohol from recent raw materials of the most recent vegetation period with at least 85 % mass ethanol as reference material. U.K.
5. APPARATUS U.K.
5.1. Multi-channel liquid scintillation spectrometer with processor and automatic external standardisation and display of the external standard/channel ratio (usual design: three meter channels and two external standard channels). U.K.
5.2. Low-potassium counter tubes suitable for the spectrometer, with dark screw-tops containing a polyethylene insert. U.K.
5.3. Volumetric pipettes, 10 ml. U.K.
5.4. Automatic dosing device 10 ml. U.K.
5.5. 250 ml round-bottom flask with ground-glass stopper. U.K.
5.6. Alcohol distillation apparatus with heating mantle, e.g. type Micko. U.K.
5.7. Microliter syringe 50 μl. U.K.
5.8. Pycnometer funnel, pycnometers, 25 ml and 50 ml. U.K.
5.9. Thermostat with a temperature stability of ± 0,01  o C. U.K.
5.10. Official alcohol tables in accordance with the Council Directive of 27 July 1976 on the approximation of the laws of Member States relating to alcohol tables, published by the Commission of the European Communities (ISBN 92-825-0146-9). U.K.
6. PROCEDURE U.K.
6.1. Adjusting the equipment U.K.

The equipment should be adjusted according to the manufacturer's instructions. Measuring conditions are optimal when the value E 2 /B, the quality index, is at its maximum.

E

=

efficiency

B

=

background

Only two meter channels are optimised. The third is left fully open for control purposes.

6.2. Selection of counter tubes U.K.

A larger number of counter tubes than will later be needed are each filled with 10 ml of 14 C-free synthesis ethanol and 10 ml toluene scintillator. Each is measured for at least × 100 minutes. Tubes whose backgrounds vary by more than ± 1 % rel. from the mean are discarded. Only tubes new from the factory and from the same batch may be used.

6.3. Determination of the external standard/channel ratio (ESCR) U.K.

During the process of setting the channels (6.1) the ESCR is determined using the appropriate computer program when the efficiency is determined. The external standard used is 137 caesium, which is already built-in by the manufacturer.

6.4. Preparation of sample U.K.

Samples having an ethanol content of at least 85 % mass and free from impurities, which absorb at wavelengths below 450 nm can be measured. The low residue of esters and aldehydes is not a problem. After the first few ml have been discarded the sample is distilled direct into the pycnometer and the alcohol content of the sample is determined by pycnometry. The values to be determined are taken from the Official Alcohol Tables.

7. MEASUREMENT OF SAMPLES USING EXTERNAL STANDARD U.K.
7.1. Slightly extinguished samples such as those described in 6.4 with an ESCR of around 1.8 may be measured via the ESCR, which provides a measure of the efficiency. U.K.
7.2. Measurement U.K.

10 ml each of the samples prepared according to 6.4 is pipettes into a selected counter tube checked for background and 10 ml of toluene scintillator is added via an automatic dosing device. The samples in the tubes are homogenised by suitable rotary movements; the liquid must not be allowed to wet the polyethylene insert in the screw-top. A tube containing 14 C-free fossil ethanol is prepared in the same way to measure the background. To check the relevant annual 14 C value a duplicate of recent ethanol from the latest vegetation period is prepared, a tube being mixed with internal standard, see 8.

The control and background samples are placed at the beginning of the measurement series, which should contain no more than 10 samples for analysis. Total measuring time per sample is at least 2 × 100 minutes, with the individual samples being measured in part stages of 100 minutes so that any equipment drift or other defect can be detected. (One cycle therefore corresponds to a measuring interval of 100 minutes per sample).

Background and control samples should be freshly prepared every four weeks.

This method requires little time and material and is particularly suitable for non-specialist laboratories processing large numbers of samples.

In the case of slightly extinguished samples (ESCR circa 1,8) the efficiency is only negligibly affected by the change in this value. If the change is within ± 5 % rel. the same efficiency can be expected. For more greatly extinguished samples, such as denatured alcohols, the efficiency can be established via the extinction correction graph. If an appropriate computer program is not available the internal standard must be used, and this gives an unambiguous result.

8. MEASURING SAMPLES USING INTERNAL STANDARD HEXADECANE 14 C U.K.
8.1. Procedure U.K.

Control and background samples (recent and fossil ethanol) and the unknown material are each measured as duplicates. One sample of the duplicate is prepared in a non-selected tube and an accurately dosed quantity (30 μl) of hexadecane 14 C is added as internal standard (added activity around 26 269  dpm/gC approximately 43 782  cBq/gC). For the sample preparation and measuring time of the other samples see 7.2, but the measuring time for the samples with the internal standard can be reduced to about five minutes by presetting at 10 5 pulses. One duplicate each of background and control samples is used per measuring series; these are placed at the beginning of the measuring series.

8.2. Handling the internal standard and counter tubes U.K.

To prevent contamination when measuring with the internal standard these must be stored and handled well away from the area where the samples for analysis are prepared and measured. After measurement the tubes checked for background may be re-used. The screw-tops and tubes containing the internal standard are disposed of.

9. EXPRESSION OF THE RESULTS U.K.
9.1. The unit of activity of a radio-active substance is the becquerel; 1 Bq = 1 decay/sec. U.K.

Indication of specific radio-activity is expressed as becquerels relative to one gram carbon = Bq/gC.

To obtain more practical results it is best to express the results in centi-bequerels = cBq/gC.

The descriptions and formulae used in the literature, based on dpm, may be retained for the time being. To obtain corresponding figures in cBq merely multiply the dpm figure by 100/60.

9.2. Expression of results with external standard U.K.

9.3. Expression of results with internal standard U.K.

9.4. Abbreviations U.K.
cpm pr

=

the mean sample count rate over the total measuring time.

cpm NE

=

the mean background pulse rate calculated in the same way.

cpm IS

=

the amount of added internal standard added (calibration radioactivity dpm).

dpm IS

=

the quantity of internal standard added (calibration radioactivity dpm).

V

=

the volume of the samples used in ml.

F

=

the content in grammes pure alcohol per ml corresponding to its concentration.

Z

=

the efficiency corresponding to the ESCR value.

1,918

=

the number of grammes alcohol per gramme carbon.

10. RELIABILITY OF THE METHOD U.K.
10.1. Repeatability (r) U.K.

r = 0,632 cBq/g C; S (r)  = ± 0,223 cBq/g C

10.2. Comparability (R) U.K.

R = 0,821 cBq/g C; S (R)  = ± 0,290 cBq/g C.]