ANNEX IU.K.

ESSENTIAL REQUIREMENTSU.K.

1. Allowable Errors U.K.

1.3.
1.3.1. Climatic environments U.K.

The manufacturer shall specify the upper temperature limit and the lower temperature limit from any of the values in Table 1 unless otherwise specified in the Annexes III to XII, and indicate whether the instrument is designed for condensing or non-condensing humidity as well as the intended location for the instrument, i.e. open or closed.

Table 1
Temperature Limits
Upper temperature limit30 °C40 °C55 °C70 °C
Lower temperature limit5 °C– 10 °C– 25 °C– 40 °C